8

Resistance measurement 
operation

The resistance measurement 

operation is set to OFF in the default

setting. To perform resistance

measurement, you need to select

the operation among AUTO, FIXED

and V/I. For AUTO and FIXED, the

B2900A Series of SMUs performs

resistance measurement by using

the current source and voltage

measure condition automatically

set by the resistance measurement

range setting. For V/I, the B2900A

Series of SMUs performs measure-

ment by using the present source/

measure condition and calculates

the resistance value.

e. Press           and set Channel 1 Limit value to 1 V

 
f.  Rotate          to select Resistance measurement operation and press          to  
  edit it. Then press           to set Resistance measurement operation to V/I

g. Press Ch1            to turn on Channel 1 Output relay.

 

h. Press           to perform a single point measurement.

3. Perform the measurement via 2-wire connection

a. Press Ch1            to turn off Channel 1 Output relay.

b. Press                   ,                     , and then press                    to display Output  
  Connection dialogue

LAB 2: Make an Accurate Low Resistance Measurement (continued)

(1) Press Limit

(2) Enter 1 V

(1) Rotate and Press

Resistance measurement operation

to select

(2) Press V/I

(1) Press Config

(2) Press Source

(3) Press Connection

Содержание B2901A

Страница 1: ...Keysight Technologies Detailed Demonstration Guide Making Resistance Measurement Using SMU Keysight B2901A 02A 11A 12A Precision Source Measure Unit Demonstration Guide...

Страница 2: ...plication software that make it easy for you to begin making productive measure ments immediately This demonstration guide shows how easily basic resistance measurement can be made using the Keysight...

Страница 3: ...ed tweezers 4 Clip the other lead of 1 k Resistor with the red gold plated tweezers Demonstration 1 Change View mode to Single View a Press repeatedly until Single View for Channel 1 is shown in the d...

Страница 4: ...o stop making measurements periodically 3 Turn off the channel output a Press Ch1 to turn off Channel 1 Output relay LAB 1 Make a Resistance Measurement continued Measurement parameters are updated pe...

Страница 5: ...Resistor with black gold plated tweezers Why is 4 wire connection required For very low resistance measurements where the residual lead resistance is comparable to the DUT resistance the 2 wire measu...

Страница 6: ...wire connection you can see the status indicator on GUI as below although no indicator can be seen on being configured to use 2 wire connection Demonstration 1 Reset the instrument a Press and then pr...

Страница 7: ...nd select and then press to configure to use 4 wire connection c Press then press to set Channel 1 V I Source Function to I Source If can t be found on the Assist keys press to change the keys d Press...

Страница 8: ...asure condition and calculates the resistance value e Press and set Channel 1 Limit value to 1 V f Rotate to select Resistance measurement operation and press to edit it Then press to set Resistance m...

Страница 9: ...connection is 1 6 Ohm The difference that is 0 6 Ohm should be the residual lead resistance on the measurement cables LAB 2 Make an Accurate Low Resistance Measurement continued 3 Press OK 2 Press 2 W...

Страница 10: ...rely on accurate measurements www keysight com go quality Keysight Technologies Inc DEKRA Certified ISO 9001 2008 Quality Management System Keysight Channel Partners www keysight com find channelpartn...

Отзывы: