Agilent M9202A Introduction

The scope of this Startup Guide is to detail the processes of receiving and installing the Agilent M9202A PXI
Express 12-bit Wideband IF Digitizer, installing the required software, and verifying basic module operation. If
you have any questions after reviewing this information, please contact your local Agilent representative or
contact us through our website at

www.agilent.com/find/assist

Related Documentation

This Startup Guide and the documentation listed below are on the CD
and at

www.agilent.com/find/M9202A

. Select

Technical Support >

Manuals

.

l

Help system for the Soft Front Panel

l

Help systems for the device drivers (IVI-C and IVI-COM, and
LabVIEW G)

l

Specifications (Data Sheet)

Follow the Startup Sequence

This Start-Up Guide is intended to lead the user through the four steps of product installation as summarized in the
diagram below. An optional fifth step shows how to perform a operational verification of the M9202A.

Step 1

: Unpack and Inspect

Step 2

: Verify Shipment

Step 3

: Install Drivers and Software

Step 4

: Install Modules

Closely follow the startup process flow in this document. Deviating from the sequence can cause

unpredictable system behavior, damage your system, and may cause personal injury.

Agilent M9202A PXI Express 12-bit Wideband IF Digitizer Startup Guide

7

Содержание Agilent Technologies PXI Express M9202A

Страница 1: ...bit Wideband IF Digitizer M9202A Startup Guide NOTICE This document contains references to Agilent Technologies Agilent s former Test and Measurement business has become Keysight Technologies For more...

Страница 2: ...or contract clause Use duplication or disclosure of Software is subject to Agilent Technologies standard commercial license terms and non DOD Departments and Agencies of the U S Government will recei...

Страница 3: ...the voltage being measured The instrument and accessories must be used in accordance with its specifications and operating instructions or the safety of the equipment may be impaired Do not exceed th...

Страница 4: ...nt means caution risk of danger You should refer to the operating instructions located in the user documen tation in all cases where the symbol is marked on the instrument This symbol indicates the ti...

Страница 5: ...anel Connectors 13 Front Panel LED 13 Step 5 Verify Operation of the M9202A Module 14 Conducting a Self Test 15 Conduct M9202A Operational check optional 16 Requirements for Verification 16 Required H...

Страница 6: ...Documentation Map 6 M9202A PXI Express 12 bit Wideband IF Digitizer Startup Guide...

Страница 7: ...elect Technical Support Manuals l Help system for the Soft Front Panel l Help systems for the device drivers IVI C and IVI COM and LabVIEW G l Specifications Data Sheet Follow the Startup Sequence Thi...

Страница 8: ...spect for damage After unpacking the module carefully inspect it for any shipping damage Report any damage to the shipping agent immediately as such damage is not covered by the warranty Warranty info...

Страница 9: ...instrument or prevent it from moving in the shipping container Styrene pellets can also cause equipment damage by generating static electricity or by lodging in fan motors If any correspondence is req...

Страница 10: ...r operation than is required for installation The amount of space listed above is required for installation 2 Note NET Framework Runtime Components are installed by default with Windows Vista Therefor...

Страница 11: ...llowing l Agilent IO Library Suite IOLS which includes the Agilent Connection Expert This software must be installed first Version 16 2 or newer of the Agilent IO Libraries Suite is required l Soft Fr...

Страница 12: ...e triangle symbol 5 Install the module into the appropriate slot of the chassis by placing the module card edges into the front mod ule guides top and bottom Ensure that the ejector handle is pushed d...

Страница 13: ...t This input is AC cou pled and 50 terminated The recommended signal level is 1 V pk pk 4 dBm The clock signal frequency should be half of the desired sampling frequency TRG OUT This MMCX female conne...

Страница 14: ...202 module is recognized 3 Select the M9202A module and check that the correct Model Name appears in the Instrument Properties 4 Select the Installed Software tab and review the configuration data as...

Страница 15: ...e with the processor controlling it When the SFP is launched it will initiate a self calibration which takes a little over 1 minute Verify that no error is shown in the lower left hand log area The im...

Страница 16: ...or 1x BNC MMCX cable 50 Coaxial 100 cm Operational Verification Procedure Do not exceed the maximum power level to the INPUT connector 0 5 W 1 Launch the soft front panel SFP The calibration must succ...

Страница 17: ...cilloscope 3 If the problem occurs during Operational Verification disconnect the input signal and note the status of the LED on the front panel it should be GREEN Reconnect the signal and select the...

Страница 18: ...ers Manufacturers Group PCIe and PCI EXPRESS are registered trademarks and or service marks of PC SIG Microsoft Windows VisualStudio VisualC VisualC and VisualBasic are either registered trademark or...