Conduct M9202A Operational check (optional)
Figure 6 - SFP showing measured waveform
If a Problem is Found
1. Verify that you have set all configuration settings as shown above.
2. Verify that the RF Generator is ON and producing the desired signals at the end of the BNC cables. This
can be done with an oscilloscope.
3. If the problem occurs during Operational Verification disconnect the input signal and note the status of the
LED on the front panel; it should be GREEN. Reconnect the signal and select the AUTO acquisition mode;
note the behavior for subsequent reference.
4. Verify that the problem is reproducible.
5. Refer to
"Return the module for service"
on page 9, concerning the details on sending the module to Agilent
for service.