Keysight 85059V Operation and Service Guide 1-5
General Information
Clarifying the Terminology of a Connector Interface
Clarifying the Terminology of a Connector Interface
In this document and in the prompts of the PNA calibration wizard, the gender
of cable connectors and adapters is referred to in terms of the center
conductor. For example, a connector or device designated as 1.0 mm –f– has a
1.0 mm female center conductor.
A connector gage is referred to in terms of the connector that it measures. For
instance, a male connector gage has a female connector on the gage so that it
can measure male devices.
Preventive Maintenance
The best techniques for maintaining the integrity of the devices in this kit
include:
— routine visual inspection
— cleaning
— proper pin depth
— proper connection techniques
All of these are described in
. Failure to detect and remove dirt or
metallic particles on a mating plane surface can degrade repeatability and
accuracy and can damage any connector mated to it. Improper connections,
resulting from pin depth values being out of specification (see
), or from bad connection techniques, can also damage these devices.