Keysight Technologies 85059A 1.0 mm Calibration and Verification Kit
3-19
Making Connections
Handling and Storage
Handling and Storage
•
Store calibration and verification devices with end caps on, in a foam–lined
storage case.
•
Never store connectors loose in a box, in a desk, or in a bench drawer. This
is the most common cause of connector damage during storage.
•
Keep connectors clean.
•
Do not touch mating plane surfaces. Natural skin oils and microscopic
particles of dirt are easily transferred to the connector interface and are
very difficult to remove.
•
Do not set connectors contact–end down on a hard surface. The plating
and the mating plane surfaces can be damaged if the interface comes in
contact with any hard surface.
•
When you are not using a connector, use plastic end caps over the mating
plane surfaces to keep them clean and protected.