1-4
Keysight Technologies 85059A 1.0 mm Calibration and Verification Kit
General Information
Achieving Specified Frequency Performance
Broadband SOLT
Calibration
As a time saver, you can use an alternate SOLT technique. You can calibrate
from DC to 110 GHz using the standards in
technique is quicker than the full calibration outlined in
but it is substantially
less
accurate.
CAUTION
The broadband SOLT technique is not recommended. It is meant to be used
only
in situations where measurement accuracy is
not
critical.
Table 1-2
Broadband SOLT
1
Frequency
Range
Calibration
Technique
“Open”
Standard
“Short”
Standard
“Load”
Standard
“Thru”
Standard
(not used for
1-port cal)
“Isolation”
Standard
DC - 110 GHz
SOLT
1
Open
(broadband)
Short 3
(broadband)
Load BB
2
Thru
Load BB
3
1. SOLT = Short, Open, Load and Thru standards
2. The broadband load is a combination of a lossy delay line plus a 50 GHz load (see