Use, Maintenance, and Care of the Devices
Gaging Connectors
When to Gage Connectors
Gage a connector at the following times:
• Prior to using a device for the first time: record the pin depth measurement so that it can be
compared with future readings. (It will serve as a good troubleshooting tool when you
suspect damage may have occurred to the device.)
• If either visual inspection or electrical performance suggests that the connector interface
may be out of typical range (due to wear or damage, for example).
• If a calibration device is used by someone else or on another system or piece of equipment.
• Initially after every 100 connections, and after that as often as experience indicates.