85058B/E
1-15
General Information
Calibration Residual Error Specifications
Table
1-8
85058E (Economy) Residual System Calibration Error Specifications
Parameter
Cal
Method
a
Frequency
Range
Residuals
PNA Data-Based
Model
b,c
PNA Polynomial
Model
b,d
Directivity
OSL
DC to
2 GHz
30
30
OSL
2 to
10 GHz
30
30
OSL
10 to
20 GHz
30
30
OSL
20 to
35 GHz
30
30
OSL
35 to
50 GHz
28
28
OSL
50 to
67 GHz
28
28
Source
Match
OSL
DC to
2 GHz
28
28
OSL
2 to
10 GHz
28
28
OSL
10 to
20 GHz
27
27
OSL
20 to
35 GHz
26
23
OSL
35 to
60 GHz
24
22
OSL
60 to
67 GHz
24
23
Reflection
Tracking
OSL
DC to
2 GHz
0.023
0.025
OSL
2 to
10 GHz
0.021
0.027
OSL
10 to
20 GHz
0.029
0.035
OSL
20 to
35 GHz
0.028
0.035
Содержание 85058B
Страница 1: ...Keysight Technologies 85058B E 1 85 mm Calibration Kits User s and Service Guide...
Страница 2: ...85058B E i...
Страница 7: ...1 General Information...
Страница 28: ...1 22 85058B E General Information When to Calibrate...
Страница 29: ...2 Specifications...
Страница 38: ...2 10 85058B E Specifications Electrical Specifications...
Страница 39: ...3 Use Maintenance and Care of the Devices...
Страница 55: ...4 Performance Verification...
Страница 58: ...4 4 85058B E Performance Verification Recertification...
Страница 59: ...5 Troubleshooting...
Страница 64: ...5 6 85058B E Troubleshooting Contacting Keysight...
Страница 65: ...6 Replaceable Parts...
Страница 71: ...85058B E 6 7 Replaceable Parts Introduction Figure 6 2 Replaceable Parts for the 85058E Economy Calibration Kit...
Страница 73: ...A Standard Definitions...