85058B/E
1-7
General Information
Achieving Specified Frequency Performance
Achieving Specified Frequency Performance
The standards in this calibration kit allow you to perform 1- or 2-port and TRM
(thru-reflect-match) calibrations.
NOTE
For best results, before beginning calibration refer to
of a Connector Interface” on page 1-20
Offset-Short Calibration Using the 85058B (Standard) Calibration Kit
Table
1-1
Offset-Short Calibration Standards
Calibration Kit Label:
PNA: 85058B Data-Based
85058BP Polynomial
PNA File Names: 85058B Data-Based.ckt
85058BP Polynomial.ckt
Frequency
Range
Calibration
Technique
“Open”
Standard
(Reflection)
“Short”
Standard
(Reflection)
“Load”
Standard
(Reflection)
“Thru”
Standard
(not used
for 1
port
“Isolation”
Standard
a
a. For best measurement results, the isolation standard should be the equivalent impedance of the DUT.
DC
35
SOLT
b
b. SOLT = Short, Open, Load and Thru standards
Open
Short 1
Load LB
Thru
Load BB
c
c. Load BB is the lowband load using a broadband load definition.
35
50 GHz
Offset
Short 2
Short 1
Short 4
Thru
Load BB
50
67 GHz
Offset
Short 3
Short 1
Short 4
Thru
Load BB
To Combine Frequency Ranges
d
d. Combinations of the calibration methods shown in the upper half of
the frequency ranges shown in the lower half of the table.
DC
35 GHz
SOLT
and
Open
Short 1
Load LB
Thru
Load BB
Offset
35
67 GHz
Offset
Short 2
Short 1
Short 4
Thru
Load BB
Short 2
DC
67 GHz
SOLT
and
Open
Short 3
Load LB
Thru
Load BB
Offset
Short 3
Short 1
Short 4
Short 2
Содержание 85058B
Страница 1: ...Keysight Technologies 85058B E 1 85 mm Calibration Kits User s and Service Guide...
Страница 2: ...85058B E i...
Страница 7: ...1 General Information...
Страница 28: ...1 22 85058B E General Information When to Calibrate...
Страница 29: ...2 Specifications...
Страница 38: ...2 10 85058B E Specifications Electrical Specifications...
Страница 39: ...3 Use Maintenance and Care of the Devices...
Страница 55: ...4 Performance Verification...
Страница 58: ...4 4 85058B E Performance Verification Recertification...
Страница 59: ...5 Troubleshooting...
Страница 64: ...5 6 85058B E Troubleshooting Contacting Keysight...
Страница 65: ...6 Replaceable Parts...
Страница 71: ...85058B E 6 7 Replaceable Parts Introduction Figure 6 2 Replaceable Parts for the 85058E Economy Calibration Kit...
Страница 73: ...A Standard Definitions...