Operating Guide
26
Keysight 8494/95/96A/B Operating and Service Manual
Procedure
1
Connect equipment as shown in Figure 2 on page 23 with the attenuator set to
0 dB attenuation.
2
Set the test oscillator to 0.3 Vrms at 1 kHz.
3
Set the SWR meter range to 2 dB (expanded) [or for the 8494A/B to 10 dB
(expanded)] and adjust its bandwidth to the center of the adjustment range.
Fine-tune the oscillator frequency to obtain maximum meter indication.
4
Set the attenuator and the SWR meter range switch as listed in
and
verify that the SWR meter indicates within the limits shown.
Table 4-2
Attenuator and SWR settings
SWR meter range
(dB)
Attenuation (dB)
Meter indication (dB)
Minimum
Actual
Maximum
8494A/B 8495A/B
8496A/B 8494A/B 8495A/B 8496A/B 8494A/B
8495A/B
8496A/B 8494A/B
8495A/B
8496A/B 8494A/B
8495A/B
8496A/B
10
2
0
0
0
—
—
Set to 0.0 Set to 0.5
—
—
10
6
1
10
10
0.40
1.40
—
—
0.60
1.60
10
12
2
20
20
0.90
0.30
—
—
1.10
0.70
10
16
3
30
30
1.35
1.25
—
—
1.65
1.75
10
[a]
22
4
40
40
1.85
0.15
—
—
2.15
0.85
12
26
5
50
50
0.35
1.10
—
—
0.65
1.90
12
32
6
60
60
0.85
0.00
—
—
1.15
1.00
12
36
[a]
7
70
70
1.30
0.90
—
—
1.70
2.10
12
[a]
42
[a]
8
—
80
1.80
–0.15
—
—
2.20
1.15
14
46
[a]
9
—
90
0.30
0.75
—
—
0.70
2.25
14
52
[a]
10
—
100
0.80
–0.30
—
—
1.20
1.30
14
56
[a]
11
—
110
1.75
0.60
—
—
1.75
2.40
[a] Adjust range by 2 dB, if needed, to obtain an on-scale indication.