 24

Chapter  3 

Specifications and Supplemental Performance Characteristics

Supplemental Performance Characteristics

where Dx is the measured value of D. It is necessary for Ze to be below 10 %.

 Rs (ESR) Measurement

Additional error Rse[%] of the Rs measurement is calculated by additional error 
Ze [%] of |Z| measurement as follows.

If Dx



 0.1:

Rse [%] = Ze / Dx

If 0.1 

 Dx 

 0.5:

Rse [%] = ( Ze

/ Dx) × 

Dx is the measured value of D and is calculated as follows.

Dx = 2 × 

 × f × Csx × Rsx,

where

f: measurement signal frequency

Csx: measured value of Cs

Rsx: measured value of Rs.

Figure 3-1 

Additional Error in |Z| measurement

D is not expressed as a percentage but as an absolute value.

1 Dx

2

+

Содержание 16034G

Страница 1: ...Keysight 16034G Test Fixture Operation Manual...

Страница 2: ......

Страница 3: ...3...

Страница 4: ...in reprints do not necessitate a new printing date The manual part number changes when extensive technical changes are incorporated January 1999 Edition 1 part number 16034 90001 April 1999 Edition 2...

Страница 5: ...DO NOT substitute parts or modify instrument To avoid the danger of introducing additional hazards do not install substitute parts or perform unauthorized modifications to the instrument Return the i...

Страница 6: ...rrants that its software and firmware designated by Keysight Technologies for use with an instrument will execute its programming instruction when property installed on that instrument Keysight Techno...

Страница 7: ...upply Off Supply In position of push button switch Out position of push button switch Frame or chassis terminal A connection to the frame chassis of the equipment which normally includes all exposed m...

Страница 8: ...8...

Страница 9: ...ng the 16034G 16 Performing Short Correction 17 Performing Open Correction 18 DUT Measurement 19 DUT measurement over 3 MHz 20 3 Specifications and Supplemental Performance Characteristics Specificati...

Страница 10: ...10 Contents...

Страница 11: ...1 Overview...

Страница 12: ...erview Product Overview Product Overview The 16034G is designed for chip type components whose sizes range from 0201 to 1306 This test fixture can take measurements of the chip type L C R Figure 1 1 P...

Страница 13: ...shipment should be as listed in Table 1 1 If the contents are incomplete if there is mechanical damage or defect notify the nearest Keysight Technologies office If the shipping container is damaged or...

Страница 14: ...nstrument s LCUR LPOT and HIGH side Contact Pin connected to a instrument s HCUR HPOT 2 Knob For lateral adjustment of HIGH side Contact Pin 3 Contact Pin Locking Screw For securing HIGH side Contact...

Страница 15: ...2 Operation This chapter describes the proper methods for open and short correction and DUT measurement...

Страница 16: ...e shows correction and measurement by the 16034G Connecting the 16034G 1 Set the cable length to 0 m in the instrument 2 Connect the 16034G directly to the UNKNOWN terminals as shown in Figure 2 1 Fig...

Страница 17: ...llows 1 Push the HIGH side contact pin s knob to the left to make firm contact with the LOW side contact pin Figure 2 2 Tighten the contact pin locking screw to secure the HIGH side contact pin Figure...

Страница 18: ...recommended that you place the DUT on the measuring stage and precisely position the HIGH side contact pin to actual DUT width 2 Tighten the contact pin locking screw to secure the HIGH side contact...

Страница 19: ...of the measuring stage and secure the contact pin with the contact pin locking screw 2 Release the LOW side contact pin with the lever and set the DUT on the measuring stage 3 Ease back slowly on the...

Страница 20: ...DUT measurement and perform measurement at 3 MHz to determine the value of the 100 SMD resistor 2 Set the measured resistance and inductance values to the instrument as load value 3 Perform load corre...

Страница 21: ...3 Specifications and Supplemental Performance Characteristics This chapter provides specifications and supplemental performance characteristics of the 16034G test fixture...

Страница 22: ...nalyzers with four terminals Applicable DUT Type Chip components Applicable DUT dimensions Maximum Voltage 40 V peak max AC DC Operating Environment temp 0 C to 55 C humidity 15 to 95 RH wet bulb temp...

Страница 23: ...ORT correction Additional errors are calculated as follows Z Measurement Additional error Ze of the Z measurement is calculated by substituting the values in the table below into the following equatio...

Страница 24: ...l error Rse of the Rs measurement is calculated by additional error Ze of Z measurement as follows If Dx 0 1 Rse Ze Dx If 0 1 Dx 0 5 Rse Ze Dx Dx is the measured value of D and is calculated as follow...

Страница 25: ...ics Supplemental Performance Characteristics Contact Pressure The following data are supplemental performance characteristics for the spring that applies contact pressure Spring constant 37 gf mm 10 S...

Страница 26: ...26 Chapter 3 Specifications and Supplemental Performance Characteristics Supplemental Performance Characteristics...

Страница 27: ...This information is subject to change without notice Keysight Technologies 1999 2001 2015 2016 Edition 6 November 2016 16034 90011 16034 90011 www keysight com...

Отзывы: