Chapter  2 

Operation

Performing Open and Short Correction

Performing Open Correction

The open correction procedure is as follows.

1. Push the HIGH side contact pin so that the distance between the HIGH and the LOW 

contact pins matches the DUT’s width (

Figure 2-3

).

It is recommended that you place the DUT on the measuring stage and precisely 
position the HIGH side contact pin to actual DUT width.

2. Tighten the contact pin locking screw to secure the HIGH side contact pin.

Figure 2-3 

Contact pin position for open correction

3. Perform the open correction as described in the specific instrument’s manual.

Before performing open correction, remove the DUT used for positioning by pulling 
back the lever to release the LOW side contact pin.

Содержание 16034G

Страница 1: ...Keysight 16034G Test Fixture Operation Manual...

Страница 2: ......

Страница 3: ...3...

Страница 4: ...in reprints do not necessitate a new printing date The manual part number changes when extensive technical changes are incorporated January 1999 Edition 1 part number 16034 90001 April 1999 Edition 2...

Страница 5: ...DO NOT substitute parts or modify instrument To avoid the danger of introducing additional hazards do not install substitute parts or perform unauthorized modifications to the instrument Return the i...

Страница 6: ...rrants that its software and firmware designated by Keysight Technologies for use with an instrument will execute its programming instruction when property installed on that instrument Keysight Techno...

Страница 7: ...upply Off Supply In position of push button switch Out position of push button switch Frame or chassis terminal A connection to the frame chassis of the equipment which normally includes all exposed m...

Страница 8: ...8...

Страница 9: ...ng the 16034G 16 Performing Short Correction 17 Performing Open Correction 18 DUT Measurement 19 DUT measurement over 3 MHz 20 3 Specifications and Supplemental Performance Characteristics Specificati...

Страница 10: ...10 Contents...

Страница 11: ...1 Overview...

Страница 12: ...erview Product Overview Product Overview The 16034G is designed for chip type components whose sizes range from 0201 to 1306 This test fixture can take measurements of the chip type L C R Figure 1 1 P...

Страница 13: ...shipment should be as listed in Table 1 1 If the contents are incomplete if there is mechanical damage or defect notify the nearest Keysight Technologies office If the shipping container is damaged or...

Страница 14: ...nstrument s LCUR LPOT and HIGH side Contact Pin connected to a instrument s HCUR HPOT 2 Knob For lateral adjustment of HIGH side Contact Pin 3 Contact Pin Locking Screw For securing HIGH side Contact...

Страница 15: ...2 Operation This chapter describes the proper methods for open and short correction and DUT measurement...

Страница 16: ...e shows correction and measurement by the 16034G Connecting the 16034G 1 Set the cable length to 0 m in the instrument 2 Connect the 16034G directly to the UNKNOWN terminals as shown in Figure 2 1 Fig...

Страница 17: ...llows 1 Push the HIGH side contact pin s knob to the left to make firm contact with the LOW side contact pin Figure 2 2 Tighten the contact pin locking screw to secure the HIGH side contact pin Figure...

Страница 18: ...recommended that you place the DUT on the measuring stage and precisely position the HIGH side contact pin to actual DUT width 2 Tighten the contact pin locking screw to secure the HIGH side contact...

Страница 19: ...of the measuring stage and secure the contact pin with the contact pin locking screw 2 Release the LOW side contact pin with the lever and set the DUT on the measuring stage 3 Ease back slowly on the...

Страница 20: ...DUT measurement and perform measurement at 3 MHz to determine the value of the 100 SMD resistor 2 Set the measured resistance and inductance values to the instrument as load value 3 Perform load corre...

Страница 21: ...3 Specifications and Supplemental Performance Characteristics This chapter provides specifications and supplemental performance characteristics of the 16034G test fixture...

Страница 22: ...nalyzers with four terminals Applicable DUT Type Chip components Applicable DUT dimensions Maximum Voltage 40 V peak max AC DC Operating Environment temp 0 C to 55 C humidity 15 to 95 RH wet bulb temp...

Страница 23: ...ORT correction Additional errors are calculated as follows Z Measurement Additional error Ze of the Z measurement is calculated by substituting the values in the table below into the following equatio...

Страница 24: ...l error Rse of the Rs measurement is calculated by additional error Ze of Z measurement as follows If Dx 0 1 Rse Ze Dx If 0 1 Dx 0 5 Rse Ze Dx Dx is the measured value of D and is calculated as follow...

Страница 25: ...ics Supplemental Performance Characteristics Contact Pressure The following data are supplemental performance characteristics for the spring that applies contact pressure Spring constant 37 gf mm 10 S...

Страница 26: ...26 Chapter 3 Specifications and Supplemental Performance Characteristics Supplemental Performance Characteristics...

Страница 27: ...This information is subject to change without notice Keysight Technologies 1999 2001 2015 2016 Edition 6 November 2016 16034 90011 16034 90011 www keysight com...

Отзывы: