8 Operation and Service Manual 11612-90001
11612A Bias Network
-
— the part number and serial number of each device
— the type of service required
— a
detailed
description of the problem and how the device was being used
when the problem occurred
Performance Tests
The procedures in this section test the electrical performance of the 11612A
using the specifications listed in this document as performance standards.
Record the results of the performance tests in the Performance Test Record,
located at the end of the procedures.
Return Loss of Input and Output Ports
1.
Connect the equipment for a standard reflection measurement.
2.
Calibrate the system with an open and a short.
3.
Connect the appropriate RF port of the 11612A to the network analyzer
test port. Terminate the opposite port with a 50 ohm load.
4.
Measure the return loss. Refer to the specifications listed in this
document.
Insertion Loss
1.
Connect the equipment for a standard insertion loss measurement.
2.
Calibrate the system with a through line.
3.
Replace the through line with the 11612A.
4.
Measure the insertion loss. Refer to the specifications listed in this
document.
The return loss of the terminating load should be at least 20 dB better than
the desired measurement value.
Table 4-1
Performance Test Record
ELECTRICAL SPECIFICATIONS
Serial Number: ______________
Tested by: ______________
Date: ______________