background image

 4     

  Operation and Service Manual 11612-90001

11612A Bias Network

-

Supplemental Characteristics

Supplemental characteristics are non-warranted performance parameters. 

They are included to provide useful operating information. 

11612A Supplemental Characteristics

11612A Option 001 Supplemental Characteristics

Pin Depth

a

0.000 to 0.127 mm (0.0000 to 0.0050 inches)

Maximum RF Input Power

+24 dBm

Impedance

50 ohms, nominal

dc

 Resistance from Bias Port to Output Port (typical)

--

At 23 

°

C

0.5 ohms

At 23 

°

C, full rated current

0.7 ohms

RF Connectors

--

RF Input

3.5 mm

b

 female

RF Output

3.5 mm male

Bias Connector (non-hermetic)

SMB male

Net Weight

0.075 kg (0.165 lb)

Dimensions (nominal, not including connectors)

 30 mm x 33 mm x 25 mm (1.2 in x 1.3 in x 1.0 in)

a. The electrical performance of the bias network is independent of its pin depth within the range stated above. 

b. 3.5 mm connectors mate with SMA connectors.

Pin Depth

a

0.000 to 0.127 mm (0.0000 to 0.0050 inches)

Maximum RF Input Power

+24 dBm

Impedance

50 ohms, nominal

dc

 Resistance from Bias Port to Output Port (typical)

--

At 23 

°

C

0.1 ohms

At 23 

°

C, full rated current

0.12 ohms

RF Connectors

--

RF Input

3.5 mm

b

 female

RF Output

3.5 mm male

Bias Connector (non-hermetic)

SMB male

Net Weight

0.075 kg (0.165 lb)

Dimensions (nominal, not including connectors)

 30 mm x 33 mm x 25 mm (1.2 in x 1.3 in x 1.0 in)

a. The electrical performance of the bias network is independent of its pin depth within the range stated above. 

b. 3.5 mm connectors mate with SMA connectors.

Содержание 11612A

Страница 1: ...Operating and Service Manual Includes Option 001 Serial Numbers This manual applies directly to Keysight 11612A Bias Networks with serial number prefix 2301A and higher Keysight 11612A Bias Network...

Страница 2: ...Software to U S government customers under its standard commercial license which is embodied in its End User License Agreement EULA a copy of which can be found at http www keysight com find sweula Th...

Страница 3: ...1 Keysight Technologies 11612A Bias Network Operation and Service Manual 11612A Bias Network Figure 1 11612A Dimensions In the following graphic all dimensions are in millimeters...

Страница 4: ...contents of this manual apply to bias networks having a serial number equal to or higher than the serial number listed on the title page of this manual Incoming Inspection If the bias network or its...

Страница 5: ...GHz 14 dB 1 to 18 GHz 18 dB 18 to 26 5 GHz 14 dB Insertion Loss Maximum 0 4 to 12 4 GHz 1 0 dB 12 4 to 26 5 GHz 1 5 dB Maximum Bias Current 2 amps dc up to 32 amps in pulsed mode Refer to Figure 5 on...

Страница 6: ...lb Dimensions nominal not including connectors 30 mm x 33 mm x 25 mm 1 2 in x 1 3 in x 1 0 in a The electrical performance of the bias network is independent of its pin depth within the range stated...

Страница 7: ...capacitance acts as a low pass filter Bias Network Schematic Diagram Figure 3 shows the 11612A bias network connected in a typical measurement setup Although other applications are possible the gener...

Страница 8: ...k so the correct bias level and repetition rate can be set before the bias network is connected Transistors and negative resistance devices may oscillate if the bias port of the 11612A bias network is...

Страница 9: ...and repetition rate are within the allowed limits remove the resistor and insert the bias network Storage and Shipment Store and ship the bias network within the following limits Returning a Device t...

Страница 10: ...for a standard reflection measurement 2 Calibrate the system with an open and a short 3 Connect the appropriate RF port of the 11612A to the network analyzer test port Terminate the opposite port with...

Страница 11: ...nput Port Input Port 0 045 to 8 GHz 0 4 to 1 GHz 8 to 18 GHz 1 to 18 GHz 18 to 26 5 GHz 18 to 26 5 GHz Output Port Output Port 0 045 to 8 GHz 0 4 to 1 GHz 8 to 18 GHz 1 to 18 GHz 18 to 26 5 GHz 18 to...

Страница 12: ...10 Operation and Service Manual 11612 90001 11612A Bias Network Figure 5 11612A Option 001 Duty Cycle Curves Overall View...

Страница 13: ...measurement needs and information on finding a local Keysight office are available on the Web at www keysight com find assist In any correspondence or telephone conversation refer to the Keysight pro...

Страница 14: ...12 Operation and Service Manual 11612 90001 11612A Bias Network...

Страница 15: ...Installation Note Xxxxx xxxxx 3...

Страница 16: ...This information is subject to change without notice Keysight Technologies 1989 2021 Edition 1 August 2021 www keysight com...

Отзывы: