Place the 400
µ
m calibration foil on the substrate and
measure it 5 times.
In some cases, depending on the material being
measured, a numerical value quite different from the
display example may be displayed. However, the
value will be automatically set to ZERO when the
ENT key is pressed for the next operation.
Use the keyboard to input the calibration foil
thickness value.
Input of calibration parameters for application No. 4
is complete.
* STD3
420
m
* STD3
412
m
•
•
•
•
* d=?
m
* d=400
m
* NF
17
Notes:
• The measurement value displayed during calibration is the
average value.
• If you input the numeric value for a standard sheet thickness
incorrectly, press the C (Clear) key to erase it and then input the
correct value. However, the values cannot be cleared if the E N T
key has already been pressed. If the ENT key has already been
pressed press the CAL key again and start the calibration procedure
over from the beginning.
400
➡
➡
Step
Key Operations
Display
Explanation
&
*
(
)
• If you have made mistakes during measurement, etc., press the
CAL key again and start the calibration procedure over from the
beginning.
• "
*
# ERROR" will appear on the LCD panel if you attempt to
make a 2-point calibration using standard sheets of the same
thickness (STD1 and STD2 for example). If this should occur, press
the CAL key again and start the calibration procedure over from the
beginning.