background image

14

7-3. Continuous measurement mode setting

If the continuous measurement mode is selected, measurements 
can be made without pressing the 

 or 

 key for each 

measurement.

* The  

 symbol denotes a blinking display.

(1)  Continuous measurement mode setting
  In the step "7-1. Measurement (4)" on page 11, pressing the 

  

key causes the decimal point to blink and display the moisture value 

and bar graph. When the bar sensor is not inserted into the sample 

or the moisture content is below the measurement range, “LO” will be 

displayed, and the numeric display will blink.

(2)  Canceling the continuous measurement mode
  Hold down the 

 key for 2 seconds or more until a “beep” 

sounds. The display will momentarily go blank. When you take your 

finger off the key, the unit will return to normal measurement mode. 

*  Pressing the 

  key turns off the power and automatically cancels 

the continuous measurement mode.

*  Note that the battery life is shorter in the continuous measurement 

mode. The battery life is approximately 48 hours when measurements 

are continuously conducted (at 20°C).

Содержание HB-600

Страница 1: ...Operating Manual Thank you for purchasing this product Please read the operating manual carefully and use this product properly Raw Cashew Moisture Tester HB 600...

Страница 2: ...given to the safety of the product however take care to read the precautions in this operating manual for proper handling Be sure to carefully follow all safety precautions Carefully read the operatin...

Страница 3: ...it Keys control panel 8 6 Before Measuring 9 7 Directions 10 7 1 Measurement 10 7 2 Average display 13 7 3 Continuous measurement mode setting 14 7 4 Moisture Content Value Correction 15 7 5 Setting t...

Страница 4: ...ment is made or key is operated for approximately 5 minutes the power is automatically shut off and avoids wasting the batteries Upper limit setting function Setting any moisture value enables a buzze...

Страница 5: ...ensation Functions Upper limit alarm setting 7 to 30 and OFF Moisture value bias adjustment 9 9 to 9 9 Auto power off automatically turned off in approx 5 minutes Average value display Continuous meas...

Страница 6: ...6 Main unit Accessories Connector Display Control panel Battery cover back Bar sensor PU 360 Operating Manual 1 5 V batteries AA alkaline x6 Shoulder strap 3 Part names...

Страница 7: ...7 Calibration curve number Battery indicator Number of measuring Calibration curve name Set value for the alarm Moisture value Bar graph 4 Display...

Страница 8: ...5 Description of Main Unit Keys control panel Key Functions Turns the power ON and OFF Adjusts a moisture value bias Selects a sample number Averages measured values Enters a minus correction value Ac...

Страница 9: ...g to secure the connection At this moment direct the silver indentation to the front side or measured values may be affected 3 Prepare a sample to be measured and allow the sample to equilibrate to th...

Страница 10: ...and Errors on page 20 2 Select the number of the calibration curve to be measured This instrument has only one pre installed calibration curve but this function is required when a user calibration cu...

Страница 11: ...er Failure to observe this may apply excessive load to the bar sensor and accordingly cause damage to it In addition the state of contact with a sample significantly changes which may greatly affect a...

Страница 12: ...step 3 Insert the bar sensor into the sample on page 11 When measurement is completed press the key to turn off the power The auto power off function turns off the power of this unit automatically if...

Страница 13: ...is continuously made measurement number will be 1 If the number of measurements exceeds 9 the measured value up to that point resets and measurement starts from measurement number 1 Due to the presen...

Страница 14: ...o the sample or the moisture content is below the measurement range LO will be displayed and the numeric display will blink 2 Canceling the continuous measurement mode Hold down the key for 2 seconds...

Страница 15: ...ias can be adjusted 9 9 to 9 9 by the following method The symbol denotes a blinking display 1 Press the key The BIAS indicator will blink and the previously entered bias value will be displayed The i...

Страница 16: ...1 Press the key The numbers to the right of ALARM will blink 2 Enter the alarm value Enter a 2 digit number To enter 20 press the and keys in succession 3 The alarm setting value is displayed Press th...

Страница 17: ...e level of sample are 1 10 slope and 1 0 intercept The user correction formula is shown as follows Measurement Value after Correction Correction Coefficient A Measurement Value before Correction Corre...

Страница 18: ...and the unit will be the calibration curve input mode 3 Enter a user calibration curve number 02 to 15 to be registered In this case 03 is selected for user calibration curve registration Successivel...

Страница 19: ...he default value is 0 0 In this case 1 0 is entered for the correction coefficient B Successively press keys in the order of 6 After the entry is completed the unit is in the state where the registere...

Страница 20: ...ure measurement Servicing is required The power is turned off after error display in 4 seconds HI always If the measurement value is always displayed as HI an abnormality may occurs in this unit Press...

Страница 21: ...21...

Страница 22: ......

Страница 23: ......

Страница 24: ...displayed on this manual may differ from the actual ones however operations and functions are not affected All efforts have been made to ensure the contents of this manual are accurate However if you...

Отзывы: