19
(2) Automatic adjustment mode
[Adjustment]Application 1 Short Brown
Ingredient values of the registered
standard sample.
Protein 0 6 . 0 %
Moisture 1 5 . 5 %
Amilose 2 0 . 0 %
Change the setting.
[Adjustment]Application 1 Short Brown
BIAS / Protein - 0 0 . 1 %
/ Moisture + 0 0 . 1 %
/ Amilose + 0 0 . 1 %
BIAS Auto-adjustment with STD. sample.
Select the item.
[ Measurement ]
[ Application ] 1 Short Brown
[ Display ]
[ Adjustment ] Serial number
[ Time ] AB0 1234
[ Communication ] VER 1 .00
2 0 0 9 - 0 5 - 3 1 2 3 : 5 9
Operation
Display
Explanation
1. By using
keys, select
the “Adjustment” setting.
2. Press the
key, and select
“Automatic adjustment” by using
keys.
3. Press the
key.
W h e n y o u w a n t t o a d j u s t
individual composition values,use
keys, and change the
highlighted composition. Then
press the
key. Using
keys, you will be able to
revise the values. When they are
to your selection, press the
key.
While you press the key, you can
increase and decrease the values
quickly with
keys.
[Adjustment]Application 1 Short Brown
Ingredient values of the registered
standard sample.
Protein 0 6 . 0 %
Moisture 1 5 . 5 %
Amilose 2 0 . 0 %
Start BIAS auto-adjustment.