Note : Sometimes, BIAS adjustment doesn’t run correctly - For example when the windows on sample
case is dirty or when the sample becomes rather damaged. If it get dirty, measure after wiping the
dirt off of it. In case that the sample is damaged, you would be replaced new standard sample.
Note : BIAS value is also acceptable with minus, instead of plus.
BIAS
= AV. Standard - AV. Measurements
- 0.5
=
6.1
-
6.6
In case of BIAS value is mius -0.5
[Adjustment]Application 1 Short Brown
BIAS / Protein - 0 0 . 1 %
/ Moisture + 0 0 . 1 %
/ Amilose + 0 0 . 1 %
BIAS Auto-adjustment with STD. sample.
Select the item.
[Adjustment]Application 1 Short Brown
BIAS / Protein - 0 0 . 2 %
/ Moisture + 0 0 . 1 %
/ Amilose + 0 0 . 1 %
BIAS Auto-adjustment with STD. sample.
Change the setting.
[Adjustment]Application 1 Short Brown
BIAS / Protein - 0 0 . 2 %
/ Moisture + 0 0 . 1 %
/ Amilose + 0 0 . 1 %
BIAS Auto-adjustment with STD. sample.
Select the item.
[ Measurement ]
[ Application ] 1 Short Brown
[ Display ]
[ Adjustment ] Serial number
[ Time ] AB0 1234
[ Communication ] VER 1 .00
2 0 0 9 - 0 5 - 3 1 2 3 : 5 9
Operation
Display
Explanation
1. By using
keys, select
the “Adjustment” setting.
2. Press the
key.
And select item by using
key.
3. Press the
key.
And change bias value by using
keys.
While you press the key, you can
increase and decrease the values
quickly with
keys.
4. Press the
key.
If you want to set another choices,
repeat operation procedure 2~4.
If you press the key, the initial
screen will be displayed.