Note : AN-820 utilizes a transmittance measurement method, empty space in the sample cases will lead
to measurement errors. It is important into the sample case firmly in such a way that there are no
empty spaces.
Note : A temperature sensing element is inserted into an opening in the bottom of the sample case in
order to measure the temperature of the sample and perform temeprature compensation. However,
in order to obtain the most precise measurement results, it is best if different temperarure between
the sample and the environment in which the unit is installed are kept to a minimum.
Note : AN-820 should be used below 35
℃
in enviroment condition in order to equip precision spectro-
scope which would be damaged in high enviroment condition.
++.+ : Over the upper limited of
Display range.
--.- : Under the lower limited of
Display range.
**
: When the caliculation is
not available.
6-2. Average Display
[Measurement] Application 1 Short Brown
Average Number of measurement times =1
Q. E. V 7 5
Protein 6 . 2 %
Protein CM @ 1 5 .0 6 . 1 %
Protein DM 7 . 2 %
Moisture 1 4 . 1 %
Amilose 2 1 . 1 %
Displaying the average.
Operation
Display
Explanation
1. W h e n p e r f o r m i n g r e p e a t e d
measurements the number of
times measured increase. (1~9)
Following the 9th measurement, the
counter will automatically reset and
start over at 1. In addition, at the
times when you press the key or
keys, the next measurement will
become the first.
2. Press the
key, the average
value based on those repeated
measurements will be displayed.
On this screen, if the
key is
pressed, a new measurement can
be performed. When the average
value is displayed, the counter resets
and the next measurement becomes
the first one. If the key is pressed,
the initial screen will be displayed.
[Measurement] Application 1 Short Brown
Number of measurement times = 1
Q. E. V
*
*
Protein + + . + %
Protein CM @ 1 5 .0
*
*
.
*
%
Protein DM
*
*
.
*
%
Moisture 1 1 . 9 %
Amilose 2 0 . 7 %
Display the average.
[Measurement] Application 1 Short Brown
Number of measurement times = 1
Q. E. V 7 4
Protein 6 . 3 %
Protein CM @ 1 5 .0 6 . 3 %
Protein DM 7 . 3 %
Moisture 1 4 . 2 %
Amilose 2 1 . 5 %
Display the average.