Review the Stored Hemoglobin Test Results

 1.insert the hemoglobin code key (pink color) into the code key slot on

the meter.

   g/dL

Hb

2.

3.

The stored results with date are displayed in order from the most recent

to the oldest via pressing the “M” button. The meter will turn off automatically

in 3 seconds  when the symbol “----” is displayed on the screen.  
If you decide to stop checking the results, stop pressing the “M”  button. The

meter will turn off automatically in 30 seconds o via pressing “S” buttonr. 

Hb

4. No test can be performed when you are checking the stored results. If you

wish to perform a new test, please turn off the meter

rst and restart the

steps described in the section

Blood Glucose/Cholesterol/ Uric Acid/

Hemoglobin Testing

】.

mg/dL

UA

2.

3.

The stored results with date are displayed in order from the most recent

to the oldest via pressing the “M” button. The meter will turn off automatically

in 3 seconds  when the symbol “----” is displayed on the screen.  
If you decide to stop checking the results, stop pressing the “M”  button. The

meter will turn off automatically in 30 seconds o via pressing “S” buttonr. 

UA

4. No test can be performed when you are checking the stored results. If you

wish to perform a new test, please turn off the meter

rst and restart the

steps described in the section

Blood Glucose/Cholesterol/ Uric Acid/

Hemoglobin Testing

】.

38

20

Содержание Multi Check

Страница 1: ...anging the Unit of Measurement 10 The Batteries Installation 10 Using Check Strip to Check the Meter 11 Control Check with Glucose Cholesterol Uric Acid Hemoglobin Control Solutions 12 When to Run a C...

Страница 2: ...Test Results 36 Review the Stored Cholesterol Test Results 37 Review the Stored Uric Acid Test Results 37 Deleting the Stored Memory of Blood Glucose Cholesterol Uric Acid Hemoglobin Test Results 39 T...

Страница 3: ...screen after 10 seconds for glucose 150 seconds for cholesterol and 20 seconds for uric acid and 10 seconds for hemoglobin The GCUHb Monitoring System is suitable for diabetes hypercholesterolemia an...

Страница 4: ...han 20 mg dL may give inaccurate cholesterol results Blood specimens containing ascorbic acid Vitamin C greater than 5mg dL acetaminophen greater than 15mg dL L Dopa greater than 1 25 mg dL Dopamine g...

Страница 5: ...id direct sunlight Be sure to follow your local regulations for proper disposal of used test strips and lancets Do not use this meter in a dry environment especially if synthetic materials are present...

Страница 6: ...Check Strip Wallet Two AAA Batteries 6 Items Need to be Purchased Separately Blood Glucose Cholesterol Uric Acid and Hemoglobin Test Strips comes with a code key Glucose Cholesterol Uric Acid Control...

Страница 7: ...a Strip handle You may touch here You may apply your blood on either side of the sample target area Contact bars Insert this end to the test strip slot on the meter Target area You must fill the whole...

Страница 8: ...can set correct time and date before you begin testing 6 When you finish setting the meter will turn off automatically in 30 seconds 1 The meter will automatically enter the setting mode after you in...

Страница 9: ...dle of the battery compartment NOTE The unit of blood glucose and cholesterol test results displayed on the screen is either mg dL or mmol L depending on which unit of measurement you have selected Th...

Страница 10: ...cover off the back of the meter and remove batteries 2 There is a unit changing switch in battery compartment Use a screwdriver to adjust the switch If you push it up the glucose cholesterol and hemog...

Страница 11: ...rip to Check the Meter You would need to use the check strip to check the meter when You use GCUHb Monitoring System for the rst time You drop the meter You think there is something wrong with the res...

Страница 12: ...Uric Acid Hemoglobin Control Solutions When to Run a Control Check with Glucose Cholesterol Uric Acid Hemoglobin Control Solutions You think there is something wrong with the glucose cholesterol uric...

Страница 13: ...ttle upside down slowly squeeze the bottle to form a small drop discard the first drop of the solution let the second drop touch the edge of the sample target area on the test strip The control soluti...

Страница 14: ...y get incorrect results Always write down the opening date of your Cholesterol control solution vial and test strip vial Both of them are good for only 2 months after rst opening or until the expirati...

Страница 15: ...ear just like the blood uric acid result 4 Hold the Cholesterol control solution bottle upside down slowly squeeze the bottle to form a small drop discard the rst drop of the solution let the second d...

Страница 16: ...measurement 4 Hold the Uric Acid control solution bottle upside down slowly squeeze the bottle to form a small drop discard the rst drop of the solution let the second drop touch the edge of the sampl...

Страница 17: ...y get incorrect results Always write down the opening date of your Hemoglobin control solution vial and test strip vial Both of them are good for only 2 months after rst opening or until the expiratio...

Страница 18: ...rip The control solution will be drawn into the reaction zone automatically 5 You will hear a Beep sound Close the cap of the control solution immediately The meter starts to count down 10 seconds the...

Страница 19: ...this lot of test strip The measured Uric Acid level using the control solution should fall within the specified ranges Expiry date YYYY MM You should fill in the opening date here 9006 Lot NO GS50810A...

Страница 20: ...ems Are your glucose cholesterol uric acid hemoglobin test strips or control solutions expired Have you ever forgotten to close your glucose cholesterol uric acid hemoglobin test strip vials or contro...

Страница 21: ...strips box Cholesterol Code Key blue color in the cholesterol test strips box Uric Acid Code Key orange color in the uric acid test strips box Puncturer Lancets Alcohol Swab not included Log Book Pre...

Страница 22: ...d lancets To avoid infection you should Never reuse the lancets Never share your puncturer with others Use an alcohol pad to wipe off blood from puncturer s tip To avoid accidental injury do not leave...

Страница 23: ...ting the top portion of the adjustable tip until the setting number lines up to the arrow Settings are based on skin type Depth 1 2 for soft or thin skin Depth 3 for average skin Depth 4 5 for thick o...

Страница 24: ...umber on the code key matches the code number on the vial of test strips you use 2 Take one strip from the vial Close the vial quickly Cautions The test strips can be damaged when they are not capped...

Страница 25: ...t down 10 seconds then show your result on the screen The meter stores your result in its memory automatically NOTE How much blood should be applied Although only a small amount of blood is needed it...

Страница 26: ...uncturer CAUTIONS To avoid accidental injury do not leave used lancet in the puncturer Always remove the used lancet immediately after each test Be sure to follow your local regulations for proper dis...

Страница 27: ...ber on the vial of test strips you use 2 Take one strip from the vial Close the vial quickly Cautions The test strips can be damaged when they are not capped and stored properly 3 Insert the test stri...

Страница 28: ...result on the screen The meter stores your result in its memory automatically NOTE How much blood should be applied Although only a small amount of blood is needed it is very important that you put en...

Страница 29: ...er rmly and pull the lancet out 14 Discard the used lancet in an appropriate container with a lid 15 Screw the adjustable tip back on the puncturer Cautions To avoid accidental injury do not leave use...

Страница 30: ...number on the code key matches the code number on the vial of test strips you use 2 Take one strip from the vial Close the vial quickly Cautions The test strips can be damaged when they are not cappe...

Страница 31: ...own 20 seconds then show your result on the screen The meter stores your result in its memory automatically NOTE How much blood should be applied Although only a small amount of blood is needed it is...

Страница 32: ...e puncturer CAUTIONS To avoid accidental injury do not leave used lancet in the puncturer Always remove the used lancet immediately after each test Be sure to follow your local regulations for proper...

Страница 33: ...er on the code key matches the code number on the vial of test strips you use 2 Take one strip from the vial Close the vial quickly Cautions The test strips can be damaged when they are not capped and...

Страница 34: ...seconds then show your result on the screen The meter stores your result in its memory automatically NOTE How much blood should be applied Although only a small amount of blood is needed it is very im...

Страница 35: ...ncturer CAUTIONS To avoid accidental injury do not leave used lancet in the puncturer Always remove the used lancet immediately after each test Be sure to follow your local regulations for proper disp...

Страница 36: ...sert the glucose code key green color into the code key slot on the meter 2 Pressing the M button the 7 day average 14 day average and 28 day average will be displayed on the screen in order 80 Glu mg...

Страница 37: ...CHOL 2 The stored results with date are displayed in order from the most recent to the oldest via pressing the M button The meter will turn off automatically in 3 seconds when the symbol is displayed...

Страница 38: ...ish to perform a new test please turn off the meter rst and restart the steps described in the section Blood Glucose Cholesterol Uric Acid Hemoglobin Testing mg dL UA 2 3 The stored results with date...

Страница 39: ...the symbol is displayed on the screen please replace batteries immediately 1 Slide battery cover off from the back of the meter 2 Remove old batteries and insert 2 new AAA size batteries 1 5V into the...

Страница 40: ...rip slot Do Not soak the meter in liquid Troubleshooting Guide SYMBOL WHAT IT MEANS ACTION This is the code number stored in the meter Make sure that this code number matches the code number on the vi...

Страница 41: ...before testing again The environmental temperature is too high to perform a test HI ashes 5 times Repeat the test in a cool place about 57 2 F 104 F 14 C 40 C You may need to wait as long as 20 minute...

Страница 42: ...displayed again call your doctor immediately pm HI KETONES Glu LO pm Glu Your blood cholesterol level is higher than 400 mg dL 10 4 mmol L Re check your blood cholesterol level If HI is displayed agai...

Страница 43: ......

Страница 44: ...ed within the period No action required A blood glucose cholesterol uric acid hemoglobin test result is stored in the memory No action required The oldest glucose cholesterol uric acid hemoglobin test...

Страница 45: ...test strip 2 New meter hasn t been code 1 Please use blood glucose cholesterol uric acid hemoglobin test strips 2 Insert the code key pm E02 The code key is damaged Contact Customer Service SYMBOL WH...

Страница 46: ...ds 20 seconds Memory Capacity 200 test results 50 test results 50 test results Operating Condition Temperature Humidity 57 2 104 F 14 40 C 85 Relative Humidity Meter Storage Condition TransportationCo...

Страница 47: ...Meter Dimension HxWxD mm 88 x 64 x 22 Display LCD display 35 x 45 mm Weight 59 grams without Batteries Life of Battery More than 1000 times Technology Used Electrode based Biosensor Model All 47...

Страница 48: ...r customer services include 1 Our Customers are entitled to free repair and repla cement of spare parts If you have problems with the product please contact local distributor agent for further informa...

Отзывы: