1
.
Insert the code key come with the glucose test strips into
the code key slot on the back of the meter.
2. Take one strip from the vial. Close the vial quickly.
3. Insert the test strip into the test strip slot on the meter. The
, then the blood
Glu
Code
6005
Cautions:
Make sure that the code number shown on the screen is the same as the code number
printed on the glucose test strip vial, otherwise you may get incorrect measurement.
4. Hold the glucose control solution bottle upside down, slowly squeeze the
bottle to form a small drop, discard the first drop of the solution, let the
second drop touch the edge of the sample target area on the test strip.
The control solution will be drawn into the reaction zone automatically.
5. You will hear a “Beep” sound. Close the cap of the control solution
immediately. The meter starts to count down 10 seconds then show
your result on the screen.
6. Check if the displayed glucose value falls within
the acceptable range shown on the test strip vial.
7. Remove the test strip from the meter and discard it.
Glu
pm
pm
mg/dL
Glu
80
symbol along with date.
Glu
pm
meter will first display code number
pm
Glu
13