![Kernel Multi Check Скачать руководство пользователя страница 16](http://html1.mh-extra.com/html/kernel/multi-check/multi-check_user-manual_1957498016.webp)
8 5
5
0
1. Insert the uric acid code key into the code key slot on the back of the meter.
2. Take one strip from the vial. Close the vial quickly.
3. Insert the test strip into the test strip slot on the meter.
The meter will first display code number
,
UA
8004
then the blood symbol along with date
.
UA
Cautions:
Make sure that the code number shown on the screen is the same as the code number
printed on the Uric Acid test strip vial, otherwise you may get incorrect measurement.
4. Hold the Uric Acid control solution bottle upside down, slowly squeeze the
bottle to form a small drop, discard
the first
drop of the solution, let the second
drop touch the edge of the sample target area on the test strip. The control
solution will be drawn into the reaction zone automatically.
5. You will hear a “Beep” sound. Close the cap of the control solution
immediately. The meter starts to count down 20 seconds then show
your result on the screen.
6. Check if the displayed uric acid value falls within the
acceptable range shown on the test strip vial.
7. Remove the test strip from the meter and discard it.
UA
8
5
5
5.6
0
20
UA
8
5
5
0
pm
UA
16
Содержание Multi Check
Страница 43: ......