OSF-4-BA-e-1510

 

12 

5.4  Dioptre adjustment and focussing 

A  special  feature  of  stereo  microscopes  is  that  they  are  fitted  with  an  optical  unit 
which has a relatively high depth of field. In order to be able get the most benefit from 
this feature, each user must synchronise the focussing mechanisms for themselves. 

The steps to do this are described in the following section. 

1. Place the object to be observed on the surface under the objective.

2. Use the objective to set the smallest possible magnification.

3. Look  through  the  righthand  eyepiece  (without  dioptre  adjustment  ring)  with  the

right eye and bring the object into focus by using the focus control dials.

4. Now set the largest magnification.

5. Once again, still only looking through the righthand eyepiece, bring the object into

focus

6. Then set the smallest possible zoom factor again.

7. Then look through the lefthand eyepiece with the left eye (with dioptre adjustment

ring) and bring the object into focus, by turning the dioptre adjustment ring left or
right at the right point.

8. In order to get the highest level of accuracy when adjusting the focus, you should

repeat steps 4-7.

9. In  this  way,  the  object  being  observed  will  be  in  focus  on  any  magnification

setting.

Содержание OSF-4

Страница 1: ...mbH Ziegelei 1 D 72336 Balingen E mail info kern sohn com Tel 49 0 7433 9933 0 Fax 49 0 7433 9933 149 Internet www kern sohn com User instructions Stereo microscope KERN OSF 4 OSF 430 OSF 431 OSF 434...

Страница 2: ......

Страница 3: ...ssembly 9 5 Operation and functionality 11 5 1 Getting started 11 5 2 Adjusting the interpupillary distance 11 5 3 Adjusting the magnification 11 5 4 Dioptre adjustment and focussing 12 5 5 Adjusting...

Страница 4: ...o the electrical parts in the base of the device must not be tampered with as in this area there is an additional risk of triggering an electric shock 1 2 Notes on the electrical system Before connect...

Страница 5: ...older must not be rotated back too far If you do this there is a risk that the microscope could tip over When the microscope is not being used you should fit the objective cap and cover the microscope...

Страница 6: ...these are highly flammable substances You must therefore keep it away from naked flames and electrical devices which can be switched on and off and only use it in well ventilated rooms However organi...

Страница 7: ...OSF 4 BA e 1510 6 2 Nomenclature Using the OSF 430 as an example...

Страница 8: ...7 OSF 4 BA e 1510 Special features of OSF 434 OSF 435 The focus dial has a coarse adjustment knob and a fine adjustment knob...

Страница 9: ...d of view mm Objective Stand Illumination OSF 430 Binocular WF 10x 20 mm 20 1x 3x mechanical 1W LED reflected light 0 21W LED transmitted light OSF 431 Binocular WF 10x 20 mm 20 2x 4x mechanical 1W LE...

Страница 10: ...epieces are already fitted in the tube You just need to remove the protective film which covers them When doing this make sure that you do not touch the optical lenses with your fingers In general you...

Страница 11: ...OSF 4 BA e 1510 10 Assembled stereo microscope...

Страница 12: ...e the gap between the two eyepieces must be re adjusted While you are looking through the eyepieces use one hand to hold the righthand or lefthand prism housing firmly By rotating outwards or inwards...

Страница 13: ...k through the righthand eyepiece without dioptre adjustment ring with the right eye and bring the object into focus by using the focus control dials 4 Now set the largest magnification 5 Once again st...

Страница 14: ...ng observed should collide Torque of the focus wheels OSF 434 OSF 435 You can adjust the torque of the focus wheels by turning the ring which is fitted along the axis of the lefthand focus wheel in a...

Страница 15: ...ed by the glasses symbol on the side They are also marked in the item description by an additional H example HSWF 10x 23 mm When fitting the eye cups make sure that the dioptre setting is not moved We...

Страница 16: ...application then it often makes sense to fit an external lighting unit to overcome this problem The lighting units which are suitable for devices of the OSF 4 series are goose neck lighting units see...

Страница 17: ...7 Features Model outfit Kern model Order number OSF 430 OSF 431 OSF 434 OSF 435 Eyepieces WF 5x 16 2 mm OZB A4101 WF 10x 20 mm OZB A4102 WF 15x 15 mm OZB A4103 WF 20x 10 mm OZB A4104 Stand mechanical...

Страница 18: ...ufficient The wrong bulb has been used The input voltage is too low Optical unit Problem Possible causes You can see two images The interpupillary distance is not set correctly The magnifications of t...

Страница 19: ...materials which you can dispose of at your local recycling centre Disposal of the storage box and device must be carried out by the operator in accordance with all national or regional laws in force...

Отзывы: