5.5.2 Linearity Adjustment of AES 220-4 and AEJ 220-4M
1. While pressing both
+
, press
to switch on the balance.
Release both keys when [
Ht00x
] appears (maintenance mode).
2. After a while, display changes to weight value (not the zero point).
Press
until
appears after displaying
.
3. While pressing
first, press
together and keep pressing both.
When
appears, release both keys.
4. While pressing
first, press
together and release both.
5.
starts blinking to indicate zero adjustment is performed
automatically.
6. When
appears, put the external weight of 100g on weigh-
ing pan. The display changes to
, put additional 120g (220g
in accumulate) on weighing pan.
7.
appears again. Remove 120g (100g remains on the pan).
The display changes to
. Remove 100g (nothing on the pan).
8. When display changes to
blinking, put the external weight on
the pan according to the table below (
~
).
9. The display changes to
and returns to normal weighing mode.
AES / AEJ – 220g
additional loaded weight
(accumulated load)
0g (0g)
50g (50g)
50g (100g)
50g (150g)
50g (200g)
20g (220g)
18
Содержание KERN AEJ
Страница 5: ...3 Electric Electronic Construction 3 1 Block Diagram Calibration Weight Unit AEJ only PRINT 5 ...
Страница 6: ...3 2 Whole Wiring Calibration Weight Unit AEJ only PRINT 6 ...
Страница 7: ...4 Troubleshooting 4 1 Troubleshooting Procedure 7 ...
Страница 11: ...4 4 Check of Electric Electronic Parts 1 Check of HT DP board 11 ...
Страница 30: ...7 Parts List 7 1 Explosion Diagram 30 ...