5.5 Linearity Adjustment
Set the verification switch to unlocked before doing the linearity adjustment (see
chapter 5.3).
Prior to verification, the verification switch must be moved into the verification
position.
5.5.1 Linearity Adjustment of AES 120-4 and AEJ 120-4M
1. While pressing both
+
, press
to switch on the balance.
Release both keys when [
Ht00x
] appears (maintenance mode).
2. After a while, display changes to weight value (not the zero point).
Press
until
appears after displaying
.
3. While pressing
first, press
together and release both.
4.
starts blinking to indicate zero adjustment is performed
automatically.
5. When display changes to
blinking, put the external weight on
the pan according to the table below (
~
).
6. When display appears
, press
key. The display changes to
and returns to normal weighing mode.
AES / AEJ – 120g
additional loaded weight
(accumulated load)
0g (0g)
30g (30g)
30g (60g)
30g (90g)
30g (120g)