13
- GAB/GAS-SH-e-0711
2.4 Press the ZERO key if the count diverges from 5000
counts.
2.5 Put the weight same as weighing capacity on the
weigh platter, then press the MODE key
→
"CAL" is
displayed, then the A/D count becomes "35000" on
the display.
If the count diverges from 5000 counts, unload the
weight and repeat the operations as set out in 2.4 and
2.5.
2.6 When the C3 mode is required, press the TARE key.
- When finishing, press the Memory Switch to
record the data to the memory, and enter the Ending
Test Mode.
•
“CAL” operation can not be performed unless zero-point adjustment is finished.
•
“CAL” operation can not be performed unless the original A/D value exceeds
33300 counts when the weight is loaded.
2.1.9
F Mode – Setting measuring conditions and EPROM clear
•
All data has been fixed according to country specifications.
•
Altering data may not conform to weighing and measuring tests for a country.
•
See 2.1.6 “C1 Mode-Country No. Setting” for Measurement Conditions Setup
Table.
1) Operation
Operation
Display
1. Stating
Test
2. Setting measuring conditions
2.1 Press the TARE key three times
2.2 Press the MODE key
→
The least significant digit blinks.
Whenever the MODE key is pressed, the required digit can
be selected.
2.3 Whenever the MODE key is pressed, the figure of the digit
which has blinked does the increment.
Содержание GAB 15K5DM
Страница 5: ...5 GAB GAS SH e 0711 1 3 Appearance Front view Rear view Bottom view...
Страница 7: ...7 GAB GAS SH e 0711 1 5 Outer Dimensions Small size Large size...
Страница 15: ...15 GAB GAS SH e 0711 Chapter 3 Hardware Configuration 3 1 Mechanisms 3 1 1 Small Size 3kg 6kg...
Страница 17: ...17 GAB GAS SH e 0711 3 1 2 Large Size 15kg 30kg...
Страница 28: ...28 GAB GAS SH e 0711 7 Start the test mode and perform the C3 mode span adjustment...