![KERN EG 220-3NM Скачать руководство пользователя страница 8](http://html1.mh-extra.com/html/kern/eg-220-3nm/eg-220-3nm_operating-instruction_1957204008.webp)
EW/EG-N/EWB-e-1627
8
KERN
EW 6200-2NM
EW 12000-1NM
Readout
0,01 g
0,1 g
Verification value (e)
0,1 g
1 g
Weighing range (max.)
6 200 g
12 000 g
Class of accuracy
I
II
Taring range (subtractive)
6 200 g
12 000 g
Minimum load (Min)
1 g
5 g
Minimum weight for counting
parts
0,01 g
0,1 g
Reference quantity
10, 30, 50, 100
10, 30, 50, 100
Reproducibility
0,02 g
0,1 g
Linearity
± 0, 03 g
± 0,1 g
Recommended adjusting
weight, not included (class)
5 kg (E2)
10 kg (F1)
Stabilization time
3 sec.
3 sec.
Weight kg (net)
2,8
2,8
Units,
verification switch in verifica-
tion position,(chap. 5.10)
g, ct
Units ,
verification switch not in veri-
fication position,
(chap. 5.10)
g, ct, oz, lb, ozt, dwt, GN, tl (HongKong), tl (Taiwan),
tl (Singapore, Malaysia), momme, tola
Air humidity
max. 80 % relative (not condensing)
Permissible ambient
condition
10° C to 30° C
Weighing plate stainless steel
180 x 160 mm
Balance dimensions
190 x 265 x 90 mm (excluding draft shield)
Vibratory filter
4
Mains supply
Mains adapter 230 V, 50/60 Hz ; 9 V DC balance,
200 mA
Interface
RS 232 C interfaced
Rechargeable battery
optional
Underfloor weighing
Hanging loop optional