![KERN EG 220-3NM Скачать руководство пользователя страница 39](http://html1.mh-extra.com/html/kern/eg-220-3nm/eg-220-3nm_operating-instruction_1957204039.webp)
EW/EG-N/EWB-e-1627
39
From now on the applied weight is
shown as a percentage.
Use the
key to return to the desired
weighing mode.
Information:
•
If the “
o-Err
“ error message appears:
-
the reference weight is outside the weighing range (see
chap. 1
”Tech-
nical Data“).
-
the set key was operated during step 2 when the weight had been ap-
plied.
•
The 100% reference is preserved until the balance is disconnected from the
mains.
6.2.4 Weighing using a tolerance range
(excl. model KERN EW120-4NM)
This balance can be used as a measuring and sorting balance, whereby the respec-
tive lower and upper tolerance limits are programmable.
Limit values can be entered in the following operating modes:
•
Weighing
•
Unit counting
•
Percent weighing
0 I I I F
85.37
%
g
Pcs
%