7.2 List of the function parameters
The balance has been set to a certain standard configuration in the factory. This is
marked by a
☆
.
Function
Display
Choice
Description of the choice
possibilities
Bar graph
1
b.G.
0 Off
☆
1 On
Tolerance weighing
(excl. EW-120-4NM)
2
SEL
☆
0 Off
1 On (chap. 7.2.1)
Zero alignment
3
A.0
0 No zero point correction
☆
1 Automatic zero point correction activated.
Automatic shutoff after 3 min.
for rechargeable battery op-
eration (function is only avail-
able inrechargeable battery
mode)
4
A.P.
0 Automatic shutoff deactivated for re-
chargeable battery operation (optional).
☆
1 Automatic shutoff activated for rechargea-
ble battery operation (optional).
Display speed
5
rE.
0 Setting for metering
1 Sensitive and fast
2
↓
☆
3
4
5 Non-sensitive but slow
Vibratory filter
6
S.d.
1 Sensitive and fast (very tranquil installation
location).
☆
2
↓
3
4 Non-sensitive but slow (very unsettled
installation location).
5 only EW-120-4NM
6 only EW-120-4NM
Interface
7
I.F.
0 Interface not active
(excl. EWB models)
☆
1 6-digit data format (excl. EW-120-4NM)
2 7-digit data format (chap. 7.2.2)
Weight unit
(only selectable, if the cali-
bration switch is not in the
calibration position see
Ch. 5.10)
81
↓
85
S.u.
1
☆
01 (g)
2
☆
14 (ct)
15 (oz)
16 (lb)
17 (ozt)
18 (dwt)
19 (grain), (excl. EWB models)
1A (tl Hong Kong)
1b (tl Singapore,Malaysia)
1C (tl Taiwan)
1d (mom)
1E (to)
3
☆
20 (Pcs)
excl. EW-120-4NM
4
☆
IF (%)
excl. EW-120-4NM
5
☆
00 Unit not set (can not be set at 81.S.u.)