7 External calibration
7.1 Calibration of non-approval models (CXB)
♦
Before calibrating, adjust the mini jumper SWA1 to “ADJ” position (see chap. 8)
♦
Observe stable environmental conditions. A warming up time of 30 minutes is
required for stabilization.
Operation
Ensure that there are no objects on the weighing plate.
0
W
a
PRE-
TARE
Press the
key and keep it pressed down whilst simultaneously operating the
key
Weight value of calibration weight is flashing on display. However, you may enter a
value of your choice via the number keys. *
3000
W
a
PRE-
TARE
CAL
Carefully place calibration weight in the centre of the weighing plate. Calibration
will be carried out automatically after dead stop control.
Remove calibration weight during count down to zero.
The balance returns automatically into weighing mode. An error message will
appear on the display should a calibration error occur or should the calibration
weight be incorrect. Turn balance off, then restart it and repeat the calibration
process.
* The calibration should be made with the recommended calibration weight (see
chap. 1 "Technical data" in the operating manual).
Note
Use the
key to exit calibration mode. The balance returns to weighing mode.
7.2 Back to the original value of calibration
(only non-approval models)
n
Press and hold
key followed by pressing
key until the scale is counting
backward to zero, then release it.
o
The scale will show the original value, which is the value setting at the factory.