11.3 Low temperature calibration
⇒
scales must be in the control room of low
temperature
n
Press
key
+
1
key
+
5
key, then press
key to enter the temperature
calibration mode
Weight
Piece Weight
Pcs
0. 0 0 0
XX
XXXXXX
low temperature(°C)
zero internal value
o
No weight on the platter, press
.
key to calibrate zero
Weight
Piece Weight
Pcs
3 0. 0 0 0
XX
XXXXXX
full capacity value
low temperature(°C)
internal value of full capacity
p
Put the full capacity weight on the platter, press
.
key to calibrate full capacity
value.
q
After scale is stable, it will reset back to zero automatically (remove the weight
mass from the platter before it resets back to zero), then the procedures
finished. Put the weight mass on the platter again to check if the display
showing the accurate weight. If it’s correct, turn off the power, then re-adjust
the mini jumper SWA1 to “LOCK” position.
11.4 Check / Delete temperature calibration data
n
Press
key
+
1
key
+
5
key
Weight
Piece Weight
Pcs
Sho t
0 0
XXXXXX
show temperature
temperature (°C)
internal value
o
Press
2
or
8
key to check the temperature calibration data.
p
Press
2
or
8
key to select the desirable data, and double-press
.
key to
delete
it.