7.3 Calibration of approval models (CXB-M)
♦
Before calibrating, adjust the mini jumper SWA1 to “ADJ” position (see chap. 8)
♦
Observe stable environmental conditions. A warming up time of 30 minutes is
required for stabilization.
Operation
Ensure that there are no objects on the weighing plate.
Ö
Press
+
1
+
5
keys
Ö
Press
key
The display shows:
0.000
M+
0
(Example)
Ö
Press
key
The display shows:
15.000
(Example)
M+
0
Weight value of calibration weight is flashing on display.
Ö
Carefully place calibration weight in the centre of the weighing plate.
Ö
Press
key
Calibration will be carried out automatically after dead stop control.
Ö
Remove calibration weight during count down to zero.
The balance returns automatically into weighing mode. An error message will
appear on the display should a calibration error occur or should the calibration
weight be incorrect. Turn balance off, then restart it and repeat the calibration
process.
194452
194452
* The calibration should be made with the recommended calibration weight (see
chap. 1 "Technical data" in the operating manual).
10
CXB-SH-e-0913