ADJ-SH-e 
1910 
 

37

 

 

 

 

 

display the value around 100.0000 

 

 

 
 

If the value>100.0000,press the PRINT 
key, until the value correct to 100.0000 

 

 

 

If the value < 100.0000, press TARE key, 
until the value correct to 100.0000. 

 

 

 

Press Mode keyto save the value, display 

SAVE…

 

 

 

 

Display 0.0000, back to the weighing 
situation. 

 

 

Note

Before linearity calibration, be sure to restore the internal weight set value to 

system default100.0000.After calibration, readjust internal calibration weight correction 
value. 

 

14 

Restoresystemsetting

 

 

Reference

: 

  If the balance setting is confusing, it will be not working properly or the   

weighing is not accurate. 

  2. If it is necessary to repeat linearity calibration and avoid the calibration 

effect be disturbed by other data setting (for example, internal calibration 
weight correction value), can restore system initial setting and then 
repeat linearity calibration. 

This operation will restore the system's initial settings, factory settings will be 

lost.

 

 

 

 

Содержание ADJ 120

Страница 1: ...ohn com Tel 49 0 7433 9933 0 Fax 49 0 7433 9933 149 Internet www kern sohn com Servicemanual Analytical balance KERN ADJ Version 1 0 09 2019 GB ADJ SA e 1910 Import r Impex Produkter AS Gamle Drammens...

Страница 2: ...onnel training 10 6 Transport and storage 10 6 1 Testing upon acceptance 10 6 2 Packaging return transport 10 7 Unpacking Setup and Commissioning 12 7 1 Installation Site Location of Use 12 7 2 Unpack...

Страница 3: ...minutes and adjusted the level 28 10 3 2 User ExternalWeightCalibration user modeC3 0 30 11 Adjust theCornerError 31 11 1 Corner Error Inspection 31 12 Adjustment of zero count and full capability co...

Страница 4: ...ication will result in a loss of the type approval After successful repair the balance will have to be reverified before it can be used again in a statutorily regulated field Detailed instructions abo...

Страница 5: ...ADJ SH e 1910 5 3 Device overview 3 1 Components...

Страница 6: ...Turn on off Exit menu CALkey Adjustment TAREkey Taring Zeroing Save setting Pos Designation 1 Glass windshield 2 Interface RS 232 3 Weighing pan 4 Bubble level 5 Display 6 Keyboard 7 Foot screws 8 Han...

Страница 7: ...ver weighing unit Selecting an application MENUkey How to save the reference Switch on off background illumination of display keep the button pressed for a long time PRINTkey Change setting in the men...

Страница 8: ...isplay Display Description Stabilitydisplay The balance is in stand by mode g Weighing unit Gram ct Weighing unit Carat lb Weighing unit Pound oz Weighing unit Ounce Pcs Application parts counting App...

Страница 9: ...pan This may damage the measuring system Impacts and overloading exceeding the stated maximum load max of the balance minus a possibly existing tare load must be strictly avoided Balance may be damage...

Страница 10: ...cautions 5 1 Pay attention to the instructions in the Operation Manual Carefully read this operation manual before setup and commissioning even if you are already familiar with KERN balances 5 2 Perso...

Страница 11: ...ADJ SH e 1910 11 Secure all parts against shifting and damage as depicted...

Страница 12: ...ainst direct draughts due to open windows and doors Avoid jarring during weighing Protect the balance against high humidity vapours and dust Do not expose the device to extreme dampness for longer per...

Страница 13: ...s adapter Operating instructions Prior to any installation and assembly works the balance must be separated from the mains supply Install the balance at the intended workplace The right place is decis...

Страница 14: ...the information on the instrument sticker matches the local mains voltage Only use KERN original mains adapter Using other makes requires consent by KERN Important Prior to commissioning check the ma...

Страница 15: ...ve reached the operating temperature see warming up time chap 1 During this warming up time the balance must be connected to the power supply mains accumulator or battery The accuracy of the balance d...

Страница 16: ...first menu item C1 showing the current setting will be displayed 2 Select menu items Press the TARE key to select the individual menu items showing the current settings one by one 3 Change settings Pr...

Страница 17: ...nu see step 4 or 5 4 How to save settings and to exit the menu Press the ON OFF key SAVE will be displayed Any changes carried out are stored by pressing the TARE key The balance returns automatically...

Страница 18: ...C3 0 No Zerot Tracking C3 1 1D Zero point Tracking C3 2 2D Zero point Tracking C3 3 3D Zero point Tracking C3 4 4D Zero point Tracking C3 5 5D Zero point Tracking C3 6 Not documented Baud rate C4 0 1...

Страница 19: ...atic internal adjustment interval only ADJ models C7 2 4 hours C7 3 6 hours C7 4 8 hours Filter C8 0 Low interference degree C8 1 Medium interference degree C8 2 High interference degree C8 3 Not docu...

Страница 20: ...environment temperature To receive accurate measuring values it is also recommended to adjust the balance periodically in weighing operation Observe stable environmental conditions A warming up time...

Страница 21: ...ADJ SH e 1910 21 9 2 Procedure...

Страница 22: ...adjustment procedure 9 3 Manual initiation of internal adjustment Menu setting C1 2 see chapter Fehler Verweisquelle konnte nicht gefunden werden Execution Press CAL button in the weighing mode The ad...

Страница 23: ...g error message will be displayed CAL NO repeat the adjustment procedure 10 Linearity calibration 10 1 Check count value Press ON OFF to shut down the scale Press CAL and the screen displays 1 Press T...

Страница 24: ...24 ADJ SH e 1910 Press ON OF the screen displays 8888888 Press TARE and the screen displays 200 0000 or 120 0000 Press TARE and the screen displays 1 112111 or 1 1612111...

Страница 25: ...E2 in the center of the weighing pan respectively read and record the count value The count value range of the balance is shown below Zero point 100g 200g 350 450 7800 7950 15250 15550 Exit the count...

Страница 26: ...26 ADJ SH e 1910 Press CAL and the screen displays 1 Press TARE and the screen displays 2 Press CAL and TARE by turns until the screen displays 8 Press ON OFF the screen displays 8888888...

Страница 27: ...ys Lnr in and Lin successively at this time the balance goes through zero point correction Put a 50 g class E2 standard weight press TARE and the screen displays Lin 100 Remove the 50g weight and load...

Страница 28: ...a few seconds the screen shows 0 0000g Linearity calibration is completed and the balance returns to normal weighing mode 10 3 External weight calibration 10 3 1 Factory External Weight Calibration Be...

Страница 29: ...ion then carefully set the adjustment weight 100 g in the centre of the weighing plate and close the windshield door When the adjustment is successfully completed you will see the CAL up indication Re...

Страница 30: ...screen displaysCAL100 Press TARE key and the screen displays CAL 200 only valid for balances whose capacity exceeds 200 grams Press TARE key until the corresponding range of the calibration point is...

Страница 31: ...nstable displayed value of the balance faulty sensor or Main PCB or interference between the sensor and its peripheral devices b The count value of the balance falls outside the specification range c...

Страница 32: ...oaded with weight on the four corners respectively Corner A Up Corner B Down Corner C Left Corner D Right C The corner error of the balance should not exceed 3D Please refer to the below table A Up C...

Страница 33: ...cation range the four corners of the balance needs to be adjusted Remove the windshield of the balance Loosen the 3 M3 cross flat head screws and remove the windshield Install the weighing pan and wei...

Страница 34: ...adjust screw at Position 1 with socket clockwise If the value is negative rotate the adjust screw at Position 1 counterclockwise Move the weight to the upper left corner G and record the displayed va...

Страница 35: ...If it is a positive value rotate the adjust screw at Position 2 with socket clockwise If the value is negative rotate the adjust screw at Position 2 counterclockwise Repeat C F procedures until the co...

Страница 36: ...be 7800 to 8000 If Full capability count value is not in this range an external resistor box must be connected at R25 Full capability count value will be within this range by the corresponding resisto...

Страница 37: ...ue to system default100 0000 After calibration readjust internal calibration weight correction value 14 Restoresystemsetting Reference If the balance setting is confusing it will be not working proper...

Страница 38: ...e 1910 Therefore be careful before operating Press ON OFF to shut down the scale Press CAL and the screen displays 1 Press TARE and the screen displays 2 Press CAL and TARE by turns until the screen d...

Страница 39: ...eted Display 1 112111 Switch on the balance Remark There are ten inner models of this type balance 200 180 120 60 H200 H180 H120 H60 600c 150c After selecting the corresponding model the balance will...

Страница 40: ...to place the new keyboard on the response positions of the upper cover while replacing the keyboard 2 Unplug and replug the data cable 3 Cannot power on the balance displays shutdown icon after power...

Отзывы: