ADJ-SH-e
1910
37
display the value around 100.0000
If the value>100.0000,press the PRINT
key, until the value correct to 100.0000
If the value < 100.0000, press TARE key,
until the value correct to 100.0000.
Press Mode keyto save the value, display
SAVE…
Display 0.0000, back to the weighing
situation.
Note
:
Before linearity calibration, be sure to restore the internal weight set value to
system default100.0000.After calibration, readjust internal calibration weight correction
value.
14
Restoresystemsetting
Reference
:
If the balance setting is confusing, it will be not working properly or the
weighing is not accurate.
2. If it is necessary to repeat linearity calibration and avoid the calibration
effect be disturbed by other data setting (for example, internal calibration
weight correction value), can restore system initial setting and then
repeat linearity calibration.
This operation will restore the system's initial settings, factory settings will be
lost.