9
Adjustment
As the acceleration value due to gravity is not the same at every location on earth, each
balance must be coordinated - in compliance with the underlying physical weighing
principle - to the existing acceleration due to gravity at its place of location (only if the
balance has not already been adjusted to the location in the factory).
This adjustment process must be carried out for the first commissioning, after each
change of location as well as in case of fluctuating environment temperature.
To receive accurate measuring values it is also recommended to adjust the balance
periodically in weighing operation.
Observe stable environmental conditions. A warming up time (see chapter 1) is required
for stabilization. Ensure that there are no objects on the weighing pan.
9.1 ADJ Models
9.1.1
Automatic adjustment using a built-in weight
Ensure menu setting
<C1-0>
see chapter Fehler! Verweisquelle konnte nicht
gefunden werden..
Automatic adjustment is performed:
having detected a change in temperature by 2°C from the previous adjustment,
a defined time interval has elapsed, selectable 2, 3, 4, 6, 8 hours
(C7-0 – C7-4, see chapter Fehler! Verweisquelle konnte nicht gefunden werden.).