ADJ-SH-e
1910
23
In case of the adjustment error (e.g. object on the plate)
the following error message will be displayed:
<CAL NO>, repeat the adjustment procedure.
10
Linearity calibration
10.1 Check count value
Press ON/OFF to shut down the scale.
Press
CAL
and the screen displays 1
Press
TARE
and the screen displays 2
Press
CAL
and
TARE
by turns until the screen
displays 8