![KERN ADJ 120 Скачать руководство пользователя страница 25](http://html1.mh-extra.com/html/kern/adj-120/adj-120_service-manual_1957153025.webp)
ADJ-SH-e
1910
25
Press
CAL
and the screen displays a zero-point count
value, such as 397
Put a 100g and a 200 g standard weight (class E2) in
the center of the weighing pan respectively, read and
record the count value.
The count value range of the balance is shown below
Zero point
100g
200g
350~450
7800~7950
15250~15550
Exit the count value inspection, by pressing
ON/OFF
twice to reboot the balance.
10.2 Procedure linear calibration
Press ON/OFF to shut down the scale.