FIGURE 4. REAR PROGRAMMING CONNECTOR, PIN
AND TEST POINT LOCATIONS
3.5.
CONNECT BOP2X-DE CHANNEL TO NETWORK
.
1. Connect a BOP 2X-DE channel to a network via the
LAN Port (see Figure 8). Use either a standard ether-
net cable if using a HUB, or a crossover LAN cable if
connecting the channel directly to a computer. The
LAN interface is not Auto-MDI-X enabled and requires
the crossover cable.
2. Apply power to BOP 2X-DE channel. The BOP 2X-DE
will beep for less than 1 second, then will be ready for
use.
3.6.
FINDING KEPCO POWER SUPPLIES ON THE
LAN
.
The PSfind utility can be downloaded from the Kepco
web site at:
www.kepcopower.com/drivers/
drivers-dl3.htm#bit802e
This utility finds all operational Kepco power supplies
connected to the LAN and then shows the MAC and IP
addresses of the models found.
To run the utility from your PC download the psfind.zip
file to your computer. Extract psfind.exe from the zip file
to a location of your choice, then double-click psfind.exe
to run the application. A separate window opens as
shown in Figure 5.Once the search is completed, all
found units are displayed in the Select a Unit window (the
MAC address appears in parentheses). If you do not see
your device in the Select a Unit window or if the icon to the
left of the selected instrument is red (not green), make
TABLE 1. REAR PROGRAMMING CONNECTOR
TERMINAL FUNCTIONS
TERMINAL
(Figure 6)
FUNCTION
1
Current Limit Reference
3
Voltage Limit Reference
15
Preamp “B”
13
2
4
11
Preamp “A”
9
5
7
6
+E
o
Limit Input
8
+
I
o
Limit Input
32
–E
o
Limit Input
34
–
I
o
Limit Input
10
Current Monitoring
20
Voltage Monitoring
12, 14
Voltage Mode Compensating Terminals
16, 18
Current Mode Compensating Terminals
17
Voltage Comparison Amp Input
19
Current Comparison Amp Input
21
On-off Controlling Logic
22
–10V Reference Voltage 4mA Max
28
+10V Reference Voltage 4mA Max
23, 25, 27,
29, 31, 33
Sensing Common
24
Circuit Breaker Control
26, 36
Turn-off Signal Output
30
Voltage Mode/Current Mode control
35
Local/Remote control
37
+5V
38
I
o
Limit Flag
39
E
o
Limit Flag
40
E
o
Mode Flag
42
I
o
Mode Flag
41
Non-isolated Turn-off Signal Input
43, 44, 45
Not Used.
46
/ON-OFF_FLAG
47, 49
Optically Isolated Turn-off Signal Input
48
ISO_GND
50
ON-OFF_CTRL
PIN 1
PIN 49
PIN 50
PIN 2
INSTALLED
(TYPICAL)
P/N 195-0121
PIN AND TEST POINT LOCATIONS
TEST POINT