2. Set the C H 2 S E L E C T O R pushbatton to release
( i n d i v i d u a l p o s i t i o n ) .
3. Connect a voltmeter c a l i b r a t o r 1 k H z ( o r 400 H z )
output to the I N P U T c o n n e c t o r © .
4. Set the R A N G E - C H 1/2 s e l e c t o r © o n your voltmeter
to the " 3 0 m V range and set the voltmeter c a l i b r a t o r
output voltage to 3 0 m V [ 5 0 m V ] .
5. A d j u s t V R 1 0 2 until the pointer s w i n g s to the f u l l s c a l e ,
6. In t u r n , set the R A N G E - C H 1 / 2 s e l e c t o r ® to the " 1 0 V
range and set the voltmeter c a l i b r a t o r output voltage
to 10V.
7. A d j u s t V R 1 0 1 u n t i l the pointer s w i n g s to the f u l l -
scale.
8. Disconnect the voltmeter c a l i b r a t o r and connect a
wideband signal generator to the C H 1 I N P U T con-
n e c t o r ® .
9. Set the signal generator frequency to 1 k H z and set
the R A N G E - C H 1/2 selector ® on voltmeter to the
" I V range.
10. A d j u s t the signal generator output level until the
pointer swings to the f u l l s c a l e .
11. Change the signal generator frequency f r o m 1 k H z to
50 k H z .
12. A d j u s t T C 1 0 1 u n t i l the pointer swings to the f u l l
scale.
13, Repeat Steps 3 through 10.
1 4 F o r channel 2 alignment, s i m i l a r y proceed w i t h
Steps 3 through 13.
C H 1
C H 2
- 1 5 -
F i g y r e
9 - Pre " Controls on Side Panel