Maintenance and care
42
Disinfection with bleaching lye
These agents contain highly aggres-
sive hypochlorites and must not be
used with aluminum rotors!
To protect the rotor 75007587 and 75007569 as far as
possible you must take the following precautions:
1. Avoid high temperatures!
The bleaching solution and the rotor should not be
warmer than ca. 25 °C.
2. Do not let the bleaching solution act longer than
absolutely necessary!
3. After disinfection, rinse the rotor thoroughly with
distilled water and allow to dry.
Decontamination
For general radioactive decontamination, use a solu-
tion of equal parts of 70% ethanol, 10% SDS and wa-
ter. Follow this with ethanol rinses, then de-ionized
water rinses, and dry with a soft absorbent cloth.
Dispose of all washing solutions in appro-
priate radioactive waste containers!
Содержание SORVALL Biofuge primo
Страница 1: ...SORVALL Biofuge primo OPERATING INSTRUCTIONS PN 00000 0...
Страница 10: ...For your safety 6 Notes...
Страница 12: ...The Biofuge primo 8 for your notes...
Страница 42: ...Operation 38 for your notes...
Страница 61: ...Appendix 57 Appendix...
Страница 63: ...Appendix 59 Braking curves 4000 3000 2000 1000 0 0 50 100 150 2 1 3 4 5 9 8 6 7 speed min 1 time s...
Страница 64: ...Appendix 60...
Страница 71: ...Appendix 67 Autoclaving protocol Date Remark Operator Signature 1 2 3 4 5 6 7 8 9 10...
Страница 72: ...Appendix 68...