Maintenance and care
41
Rotor and rotor chamber must be treated with a neu-
tral, universal disinfectant. Best suited for this purpose
are disinfectant sprays, ensuring that all rotor and ac-
cessory surfaces are covered evenly.
•
Please use 70% ethanol for disinfection.
Please note the safety measures and
handling hints when applying these
substances!
For other disinfectants please consult KENDRO
Services!
•
You may disinfect the rotor and the accessories as
described in the following section. Be sure to follow
the pertinent safety procedures for handling infec-
tious material.
1. Pull mains plug.
2. Unscrew the rotor chuck.
3. Grab the rotor with both hands and pull it perpen-
dicularly off the drive shaft.
4. Remove the centrifuge tubes and adapters, and
disinfect them or dispose of them as necessary.
5. Treat the rotor and the rotor lid according to the
instructions given for the disinfectant (soaking in
liquid or spraying). You must strictly observe the
specified action times!
6. Turn the rotor head down and drain off the disinfec-
tant. Thereafter thoroughly rinse rotor and lid with
water.
7. Dispose of the disinfectant according to valid regu-
lations.
8. Aluminum rotors have to be treated with anticorro-
sive protective oil subsequently.
Содержание SORVALL Biofuge primo
Страница 1: ...SORVALL Biofuge primo OPERATING INSTRUCTIONS PN 00000 0...
Страница 10: ...For your safety 6 Notes...
Страница 12: ...The Biofuge primo 8 for your notes...
Страница 42: ...Operation 38 for your notes...
Страница 61: ...Appendix 57 Appendix...
Страница 63: ...Appendix 59 Braking curves 4000 3000 2000 1000 0 0 50 100 150 2 1 3 4 5 9 8 6 7 speed min 1 time s...
Страница 64: ...Appendix 60...
Страница 71: ...Appendix 67 Autoclaving protocol Date Remark Operator Signature 1 2 3 4 5 6 7 8 9 10...
Страница 72: ...Appendix 68...