INSTALLATION INSTRUCTIONS

CSMRI_F-3732_050813

GENERAL
•  Read instructions provided with the 

MedMaster™ MRIPS-312 external power 
supply for proper installation and electrical 
connection to the lighting system.

INSTALLATION INSTRUCTIONS
•  The CleanScene MRI™ lighting system 

contains sensitive electronics. Take care 
to avoid damage by improper handling or 
static electricity discharge (ESD). Either  
type of damage could render the  
system inoperable.

Lightbox Mounting Procedure
CleanScene MRI™ lightboxes are designed 
for either 1” or 1½” (G10 or G15 
respectively—see product label) T-bar grid 
systems with vertical grid T thickness of less 
than ¼”. Grid systems can be installed with 
standard 24” x 24”, 24” x 48” or 48” x 48” 
centerline grid spacing.

Use the following procedure to install the 
lightbox(es) into the ceiling grid:

•  Unpack products from the shipping carton.
•  Determine where lightbox(es) will be 

mounted and note the desired orientation 

of graphic images within room.

NOTE: For installations with a multiple-

configuration (the image is divided among 

multiple lightboxes), make sure lightbox is 

oriented toproduce a continuous image.

•  It is recommended that the door remain 

attached to the lightbox housing during 

installation. This will prevent damage to 

internal components, particularly the LED 

light sources. All electrical connections are 

accessible through the external junction 

box located on the top of the  

lightbox housing.

GENERAL
•  Lift lightbox(es) into ceiling structure using 

a recommended minimum of two able-
bodied individuals.

•  Lay lightbox(es) into grid.

•  Attach support wires between hanger 

brackets and ceiling support structure.

•  Mount and wire the MRIPS-312 external 

power supply per the procedures provided 
in the supplementar conduit and DC wiring 
to an MRI room EMI filter sized to the  
load of the power supply (supplied by 
others).completely enclosed in grounded 
aluminum conduit. Any gaps, regardless 
of size, must be closed and/or wrapped in 
copper foil tape.

•   If a 0-10V dimmer is to be connected, 

install at this time. Dimmer must be 
installed outside the shielded MRI 
environment with an intermediate MRI 
room EMI filter to the luminaire(s). Kenall 
recommends using a Leviton

®  

IllumaTech

™ 

IP710 series dimmer or Grafik Eye GRX-TVI 
control interface. Contact Kenall for 
suitability of using an alternate sink-type 
0-10V dimmer. 

Содержание CSMRI Series

Страница 1: ...OUNTING RESTRICTIONS SAVE THESE INSTRUCTIONS INSTALLATION INSTRUCTIONS PREP INSTALLATION SURFACE Disconnect power to all circuits before wiring system Install in accordance with all national state and...

Страница 2: ...age is divided among multiple lightboxes make sure lightbox is oriented toproduce a continuous image It is recommended that the door remain attached to the lightbox housing during installation This wi...

Страница 3: ...supply from the EMI filter to luminaires s using at least 18 AWG wires Class 1 wiring is required Make sure wiring is completely enclosed in grounded aluminum conduit Any gaps regardless of size must...

Страница 4: ...MRIPS 312 instruction sheet and fax to Kenall at 847 360 1781 II Door Opening Closing and Removal Procedure NOTE OPENING THE DOOR DURING NORMAL INSTALLATION IS NOT NECESSARY ALL WIRING SHOULD BE ACCE...

Страница 5: ...for proper engagement of catch hooks before closing door Installing the Door CSMRI22 Lift door by spring clips on one side Engage spring clips on one side to hang vertically Closing the Door With catc...

Страница 6: ...on opposite sides of each other and when installed into the lightbox will put the image in the correct orientation Fasten all hex nuts to retainer brackets Torque all fasteners equally to 12 lb in Re...

Страница 7: ...ed installed in applications for which it is not intended This includes operation in ambient temperatures above stated limits for any length of time Failure by electrical surge shall not be covered un...

Страница 8: ...to an ungrounded supply Wiring connections must be made internal to the enclosure Failure to install unit on a mechanically sound surface may result in personal injury physical damage or potential fi...

Страница 9: ...nduit Installation has no ungrounded unshielded portions of conduit or openings of any size or shape Check here if dimming is not applicable MRI Room EMI filters for the 24VDC supply are of the type i...

Отзывы: