background image

28

When properly purged, the hand pump will generate positive pressure quickly. The 

construction of the hand pump means that the same pressure is generated for the test 

item as for the reference test gauge.

The pressure is set to the necessary value using the fine adjustment valve. With the 

pressure relief valve, the pressure can be relieved into the reservoir. 

For the assessment of the test sample, the pressure displayed on the reference  

manometer is compared with the measurement value.

3.1  Important notes on pressure fluctuations

It is completely normal that the pressure is not constant from the start. In the event 

of changes to the pressure in the measurement device, it always lasts a few minutes 

until the pressure stabilises. 

This is influenced by a variety of factors. The most important influence factors are: 

●  Improper purging: 

  If there is still air in the measurement device, the pressure build-up lasts signi- 

ficantly longer. Moreover, there is a fall-off in pressure over a limited period due to 

diffusion processes. 

●  Mechanical characteristics of the pressure hose:

  Bending or rolling up the pressure hose causes a reduction in the volume and thus 

leads to an increase in pressure. 

  At high pressure, the pressure hose extends. Furthermore, enclosed air can be 

diffused through the pressure hose. In both cases, a drop in pressure results. 

●  Temperature influence:

  Temperature changes lead to a change in volume in the measurement device and 

therefore to changes in pressure. The smaller the volume is, the greater the change 

in pressure.

●  Settling times of reference and test sample:

  Observe the required waiting times after the reference and test sample have been 

switched on. More information on this can be found in the corresponding operating 

manual.

3  Construction and function

Содержание HTP1

Страница 1: ...21 Hydraulic Hand Pump HTP1...

Страница 2: ...ctions 26 3 Construction and function 27 3 1 Important notes on pressure fluctuations 28 4 Purging hydraulic lines 29 5 Handling 30 5 1 Generating pressure 31 5 2 Pressure measurements 33 5 3 Releasin...

Страница 3: ...spe cified order Thoroughly read and understand the information in the section Safety instructions If you have any problems or questions please contact your supplier or contact us directly at KELLER...

Страница 4: ...nal ac cessories Transportation case The transportation case provides optimal protection for the hand pump and other accessories with its tight fitting rigid foam inlay A document compartment is locat...

Страница 5: ...h 9 Technical data CAUTION Risk of injury or material damage By connecting the HTP1 the measuring device is hydraulically connected to the hand pump If the pressure relief valve is opened a compressed...

Страница 6: ...must be aware of this operating manual and have access to it at all times General safety instructions In all work the existing national regulations for accident prevention and safety in the workplace...

Страница 7: ...mp body Reservoir with suction and relief outlets Piston rod with internal spring Pressure hose with Minimess connection Function The reference and the test sample or the pressure device to be tested...

Страница 8: ...portant influence factors are Improper purging If there is still air in the measurement device the pressure build up lasts signi ficantly longer Moreover there is a fall off in pressure over a limited...

Страница 9: ...ed hydraulic fluid Open the pressure relief valve by turning counter clockwise Turn the reference manometer with matching seal in the hand pump s connection G1 4 IMPORTANT Do not tighten the reference...

Страница 10: ...ping procedure ensure that fingers or other body parts are kept away from the area between the hand grips and the piston rod CAUTION Material damage to valve stop If put under too much strain the stop...

Страница 11: ...cessary testing pressure Following this carefully increase the pressure using the fine adjustment valve Notes on operation Please observe the following notes concerning the operation of the hand pump...

Страница 12: ...clock wise direction until the stop is reached Create pressure Push the hand grips together The pressure is built up Repeat the pumping movements until the necessary testing pressure has roughly been...

Страница 13: ...ssure it takes a few minutes for the pressure in the measure ment device to stabilize paragraph 3 1 Wait approximately 3 5 min before you begin taking measurements The necessary procedures for measuri...

Страница 14: ...re is no longer any positive pressure Remove the test item with adapters and seals from the pressure hose Put the hand pump away along with any accessories which have been used paragraph 7 Maintenance...

Страница 15: ...eck seal size material wear Seal incorrectly positioned Check positioning of seal Pressure relief valve open Close pressure relief valve No further increase of pres sure is possible Air was sucked in...

Страница 16: ...rtation case which is avai lable as an optional accessory The tight fitting rigid foam inlay offers optimum protection for the hand pump with pres sure hose and accessories Reference gauge The common...

Страница 17: ...l The hand pump consists of various different materials It should not be disposed of with household waste Take the hand pump to your local recycling plant or send the hand pump back to your supplier o...

Страница 18: ...e 700 bar Medium Hydraulic oil Demineralized water optional Connection Reference G1 4 Pressure hose Minimess 1620 Test item Pressure hose 1 m with Minimess 1620 or with Minimess adapter 1620 to G1 4 D...

Страница 19: ...Transportation case Lid Burl foam with document compartment Main compartment Tight fitting rigid foam inlay for pump and accessories Dimensions 450 x 370 x 140 mm Weight with pump and accessories 4 35...