27
Special safety instructions
Warnings that are specifically relevant to individual operating procedures or activities
can be found at the beginning of the relevant sections of this operating manual.
Connections
➀
Minimess
®
connection
pressure hose
➁
Connector reference gauge
Operating elements
➂
Fine adjustment valve (hand wheel)
➃
Pressure relief valve (rotary knob)
➄
Seal reservoir with safety valve
➅
Hand grips
Main components
➆
Upper and lower pump body
➇
Reservoir with suction and
relief outlets
➈
Piston rod with
internal spring
➉
Pressure hose with
Minimess
®
connection
Function
The reference and the test sample or the pressure device to be tested (hereafter “test
sample”) are connected to the hand pump.
Pumping is carried out by repeatedly pushing the hand grips together. An internal
spring returns the hand grips to their start position.
The pumping movements are transferred to the pistons in the pump body via the push
rods. The hydraulic liquid is then suctioned from the storage container and pressed
through valves to the test item.
2 Safety instructions
3 Construction and function
➀
➁
➂
➃
➄
➅
➆
➇
➈
➉
➆
➀
➀
Содержание HTP1
Страница 1: ...21 Hydraulic Hand Pump HTP1...