Operating instructions CellaTemp PX 4x 5x 6x
PX 44
Temperature range
750 - 2400 °C
850 - 3000 °C
Sensor
Fotodiode
Spectral sensitivity
0�95 / 1�05 µm
Focussing range
0.4 m - ∞ optic PZ 20.01
Distance to target-size ratio
150:1 (PZ 20.01)
Resolution current output
0�2 K + 0�03 % of selected range
Resolution temperature
reading
1 K
Response time t
98
Temperature range 750 – 2400 °C
≤ 10 ms (T > 950 °C)
Temperature range 850 -3000 °C
≤ 10 ms (T > 1050 °C)
Measurement uncertainty
Temperature range 750 - 1500 °C
0�35 % of reading [°C] + 2 K
Temperature range 1500 - 2400 °C
0�5 % of reading [°C] + 2 K
Temperature range 2400 - 3000 °C
1 % of reading
(at ɛ =1.0 and Tu = 23 °C)
Repeatability
2 K
Temperature coefficient with
reference to 23°C
≤ 0,05 %/K
of temp� reading / K