Quick start guide CellaTemp PA
Electrical connection
The pyrometer is supplied with low voltage 24V DC (18 ... 32V DC).
The pyrometer may only be installed by a skilled, qualified
electrician. Do not connect the instrument while the voltage
supply source is turned on. Please observe international
safety regulations at all times
►
Switch to neutral and verify absence of voltage
►
Connect the instrument according to the following schematic.
PIN 1
WH (white)
L+ (Power supply 24V DC)
PIN 2
BN (brown)
Digital in- / out 1
PIN 3
GN (green)
Analogue output 2/ analogue input
PIN 4
YE (yellow)
Analogue output 1
PIN 5
GY (grey)
Serial interface RS 485 (A) /Tx-/D-)
PIN 6
PK (pink)
Serial interface RS 485 (B) /Tx+/D+)
PIN 7
BU (blue)
Digital in- / output 2
PIN 8
RD (red)
L- (GND)
The pyrometer must be protected against high voltage and strong
electromagnetic fields. Use a shielded cable. The shield must be
connected with the connector housing.
Use a flyback diode when switching inductive loads.
Installation
Install the pyrometer in a location where it will not be unnecessarily ex-
posed to smoke, ambient heat or water vapour.
1. Select a suitable location for the pyrometer. The following criteria
must be observed:
EN
Quick start guide CellaTemp PA
General
This guide gives you the minimum information to properly install the py-
rometers of the PA series. For detailed information please refer to the user
guide CellaTemp PA. You can download it from the following link:
www.keller.de/its/
Explanation of symbols
Important safety-related references in this manual are marked with a sym-
bol.
This symbol points out guidelines. If you do not observe
them, the device might be damaged, malfunctioning or even
fail to operate.
CAUTION: This symbol points out hints and information which
should be heeded for efficient and trouble-free operation.
►
Action: This symbol instructs the operator to take action.
> Reaction, Result: This symbol indicates the result of the action
taken.
Laser safety instructions
Class 2 Laser Product
●
Never look directly into the laser beam path (emitted power <1.0
mW at a wavelength of 630-670 nm)
●
Do not leave the instrument unattended when the laser is activated.
●
Do not point the laser beam at any person.
●
During pyrometer installation and alignment, make sure to avoid
the possibility of laser light reflections caused by reflective sur-
faces.
●
All currently valid laser safety standards must be observed.
Laser Warning Label
The black and yellow laser warning label is affixed next to the nameplate
of the instrument. An arrow indicates the laser emission path (lens open-
ing).
EN
Содержание CellaTemp PA 4 Series
Страница 1: ...CellaTemp PA 4x 5x 6x Ident Nr 1105004 11 2021 EN FR IT ES RU CN KR DE...
Страница 2: ......
Страница 28: ...CellaTemp PA CellaTemp PA www keller de its 2 1 0 630 680 RU...
Страница 30: ...CellaTemp PA 1 0 65 C 65 C 2 3 90 C 45 4 RU...
Страница 31: ...CellaTemp PA 5 6 Mode 3 7 RU...
Страница 32: ...CellaTemp PA Lambda 1 Lambda 2 8 9 RU CellaTemp PA 1 0 65 C 65 C 2 3 90 C 45 4 RU...
Страница 33: ...CellaTemp PA KELLER HCW RU...
Страница 34: ...CellaTemp PA PA CellaTemp PA www keller de its 1 0 mW 630 670 nm CN...
Страница 36: ...CellaTemp PA 0 65 C 65 C 11 12 90 45 13 14 15 3 MODE 2 16 CN...
Страница 37: ...CellaTemp PA Lambda 1 Lambda 2 17 18 KELLER HCW CN...
Страница 38: ...CellaTemp PA PA CellaTemp PA www keller de its Class 2 630 670 nm 1 0 mW KR...
Страница 40: ...CellaTemp PA 19 0 65 C 65 C 20 21 90 45 22 2 23 24 KR...
Страница 41: ...CellaTemp PA MODE 3 MODE 2 25 2 Lambda 1 Lambda 2 26 27 KR...
Страница 42: ...CellaTemp PA KELLER HCW KR...
Страница 43: ......
Страница 44: ...Carl Keller Stra e 2 10 D 49479 Ibbenb ren Laggenbeck Germany www keller de its...