Series 2600B System SourceMeter® Instrument Reference Manual
Section 7: TSP command reference
2600BS-901-01 Rev. B / May 2013
7-207
Details
This attribute sets the time interval between measurements when
smu
X
.measure.count
is set to a value
greater than 1. The SMU attempts to start each measurement when scheduled. If the SMU cannot keep up with
the interval setting, measurements are made as quickly as possible.
If filtered measurements are being made, the time interval is from the start of the first measurement for the
filtered reading to the first measurement for a subsequent filtered reading. Extra measurements made to satisfy a
filtered reading are not paced by this interval.
Example
smua.measure.interval = 0.5
Sets the measure interval for SMU channel A to 0.5.
Also see
reset()
(on page 7-160)
setup.recall()
(on page 7-180)
smuX.measure.count
(on page 7-201)
smuX.reset()
(on page 7-216)
smuX.measure.lowrangeY
This attribute sets the lowest measurement range that is used when the instrument is autoranging.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Instrument reset
SMU reset
Recall setup
Saved setup
Current:
Models 2601B/2602B/2604B/
2611B/2612B/2614B:
100e-9 (100 nA)
Model 2634B:
1e-9 (1 nA)
Models 2635B/2636B:
100e-12 (100 pA)
Voltage:
Models 2601B/2602B/2604B:
100e-3 (100 mV)
Models 2611B/2612B/2614B/
2634B/2635B/2636B:
200e-3 (200 mV)
Usage
lowRange
= smu
X
.measure.lowrange
Y
smu
X
.measure.lowrange
Y
=
lowRange
lowRange
The lowest voltage or current measurement range used during autoranging
X
Source-measure unit (SMU) channel (for example,
smua.measure.lowrangev
applies to SMU channel A)
Y
SMU measure function (
v
= voltage,
i
= current)
Details
This attribute is used with autoranging to put a lower bound on the range used. Since lower ranges generally
require greater settling times, setting a lowest range limit might make measurements require less settling time.
If the instrument is set to autorange and it is on a range lower than the one specified, the range is changed to the
lowRange
range value.
Example
smua.measure.lowrangev = 1
Sets voltage low range for SMU channel A to 1 V.