2-18
Calibration
• Note and record the multimeter reading.
• Send the source and sense commands using the multimeter reading as the parameter.
For example:
:CAL:PROT:SOUR 1.997
:CAL:PROT:SENS 1.997
• Send the SOUR:VOLT 0.0 command to set the source voltage to 0V.
• Note and record the multimeter reading.
• Send the :CAL:PROT:SOUR command using the multimeter reading as the command
parameter. For example:
:CAL:PROT:SOUR -1.02E-3
Table 2-8
Voltage calibration initialization commands
* Remote sensing may be used if desired, but is not essential when using the recommended
digital multimeter. To use remote sensing, send :SYST:RSEN ON.
Command
Description
*RST
Restore GPIB defaults.
:SOUR:FUNC VOLT
Activate voltage source.
:SENS:CURR:PROT 0.1
Current limit when voltage source is active.
:SENS:CURR:RANG 0.1
Make sure 1A range is not active.
:SOUR:VOLT:PROT MAX
Maximum allowable source voltage.
:SYST:RSEN OFF
Disable remote sensing. *
:CAL:PROT:CODE 'KI002420'
Unlock cal.
:OUTP:STAT ON
Turn source on.
Содержание SourceMeter 3A 2420
Страница 2: ...Model 24203A SourceMeter Service Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E...
Страница 15: ...1 Performance Verification...
Страница 31: ...2 Calibration...
Страница 53: ...3 Routine Maintenance...
Страница 56: ...3 4 Routine Maintenance...
Страница 57: ...4 Troubleshooting...
Страница 70: ...4 14 Troubleshooting...
Страница 71: ...5 Disassembly...
Страница 78: ...5 8 Disassembly...
Страница 79: ......
Страница 80: ......
Страница 81: ......
Страница 82: ......
Страница 83: ......
Страница 84: ...6 Replaceable Parts...
Страница 102: ......
Страница 103: ......
Страница 104: ......
Страница 105: ......
Страница 106: ......
Страница 107: ...A Specifications...
Страница 112: ...A 6 Specifications...
Страница 113: ...B Command Reference...
Страница 124: ...B 12 Command Reference...
Страница 125: ...C Calibration Programs...