Table 2-6
:CALibration:PROTected:SENSe parameter ranges
Sense range
First parameter
(zero)
Second parameter
(negative full scale)
Third parameter
(positive full scale)
0.2V
2V
20V
200V
1µA
10µA
100µA
1mA
10mA
100mA
1A
-0.002 to +0.002
-0.02 to +0.02
-0.2
to +0.2
-2
to +2
-1E-8 to +1E-8
-1E-7 to +1E-7
-1E-6 to +1E-6
-1E-5 to +1E-5
-1E-4 to +1E-4
-1E-3 to +1E-3
-1E-2 to +1E-2
-0.18 to -0.22
-1.8
to -2.2
-18
to -22
-180 to -220
-0.9E-6 to -1.1E-6
-9E-6
to -11E-6
-90E-6 to -110E-6
-0.9E-3 to -1.1E-3
-9E-3
to -11E-3
-90E-3 to -110E-3
-0.9
to -1.1
+0.18 to +0.22
+1.8
to +2.2
+18
to +22
+180 to +220
+0.9E-6 to +1.1E-6
+9E-6
to +11E-6
+90E-6 to +110E-6
+0.9E-3 to +1.1E-3
+9E-3
to +11E-3
+90E-3 to +110E-3
+0.9
to +1.1
NOTE: Parameter steps for each range may be performed in any order, but all three parameter steps for each
range must be completed.
Table 2-7
:CALibration:PROTected:SOURce parameter ranges
Source range
First parameter
(negative full scale)
Second parameter
(negative zero)
Third parameter
(positive full scale)
Fourth parameter
(positive zero)
0.2V
2V
20V
200V
1µA
10µA
100µA
1mA
10mA
100mA
1A
-0.18 to -0.22
-1.8
to -2.2
-18
to -22
-180 to -220
-0.9E-6 to -1.1E-6
-9E-6
to -11E-6
-90E-6 to -110E-6
-0.9E-3 to -1.1E-3
-9E-3
to -11E-3
-90E-3 to -110E-3
-0.9
to -1.1
-0.002 to +0.002
-0.02 to +0.02
-0.2
to +0.2
-2
to +2
-1E-8 to +1E-8
-1E-7 to +1E-7
-1E-6 to +1E-6
-1E-5 to +1E-5
-1E-4 to +1E-4
-1E-3 to +1E-3
-1E-2 to +1E-2
+0.18 to +0.22
+1.8
to +2.2
+18
to +22
+180 to +220
+0.9E-6 to +1.1E-6
+9E-6
to +11E-6
+90E-6 to +110E-6
+0.9E-3 to +1.1E-3
+9E-3
to +11E-3
+90E-3 to +110E-3
+0.9
to +1.1
-0.002 to +0.002
-0.02 to +0.02
-0.2
to +0.2
-2
to +2
-1E-8 to +1E-8
-1E-7 to +1E-7
-1E-6 to +1E-6
-1E-5 to +1E-5
-1E-4 to +1E-4
-1E-3 to +1E-3
-1E-2 to +1E-2
NOTE: Parameter steps for each range may be performed in any order, but all four parameter steps for each range must be
completed.
2-17
Model 2400 Service Manual
Calibration
2400-902-01 (G - Feb 2006)
Содержание SourceMeter 2400
Страница 10: ......
Страница 13: ...1 PerformanceVerification...
Страница 28: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 1 16 Performance Verification...
Страница 29: ...2 Calibration...
Страница 51: ...3 RoutineMaintenance...
Страница 54: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 3 4 Routine Maintenance...
Страница 55: ...4 Troubleshooting...
Страница 68: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 4 14 Troubleshooting...
Страница 69: ...5 Disassembly...
Страница 76: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 5 8 Disassembly...
Страница 77: ...6 ReplaceableParts...
Страница 91: ...A Specifications...
Страница 111: ...B CommandReference...
Страница 123: ...C CalibrationPrograms...
Страница 130: ...C 8 Calibration Programs...
Страница 134: ......
Страница 136: ......