SPECIFICATIONS
SPEC-2425
A 24053
Released
SZ 4/18/00
B 23093
Revised
SZ 4/28/00
C 27953
Revised
SZ 3/20/03
SJ 4/18/00
CC 4/18/00
HW 3/20/03
Rev. C
Page 1 of 3
2425 and 2425-C SourceMeter
®
Specifications
SOURCE SPECIFICATIONS
VOLTAGE PROGRAMMING ACCURACY (LOCAL OR REMOTE SENSE)
RANGE
PROGRAMMING
RESOLUTION
ACCURACY (1 Year)
23°C
±
5°C
±
(% rdg. + volts)
NOISE
(peak-peak)
0.1Hz – 10Hz
200.00 mV
5 µV
0.02% + 600 µV
10 µV
2.00000 V
50 µV
0.02% + 600 µV
50 µV
20.0000 V
500 µV
0.02% + 2.4 mV
500 µV
100.000 V
2.5 mV
0.02% + 12 mV
2.5 mV
TEMPERATURE COEFFICIENT (0°-18°C & 28°-50°C):
±
(0.15 × accuracy specification)/
°C.
MAX. OUTPUT POWER:
110W, four quadrant source or sink operation.
SOURCE/SINK LIMITS:
±
21V @
±
3.15A,
±
105V @
±
1.05A.
VOLTAGE REGULATION: Line:
0.01% of range.
Load:
0.01% of range + 100
µV
.
NOISE 10Hz – 1MHz (p-p):
50mV typical into a resistive load.
OVERVOLTAGE PROTECTION:
User selectable values, 5% tolerance. Factory default = none.
CURRENT LIMIT:
Bipolar current limit (compliance) set with single value. Min. 0.1% of range.
OVERSHOOT:
<0.1% typical (full scale step, resistive load, 10mA range).
CURRENT PROGRAMMING ACCURACY (LOCAL OR REMOTE SENSE)
RANGE
PROGRAMMING
RESOLUTION
ACCURACY (1 Year)
2
23°C
±
5°C
±
(% rdg. + amps)
NOISE
(peak-peak)
0.1Hz – 10Hz
10.0000 µA
500 pA
0.033% + 2 nA
50 pA
100.000 µA
5 nA
0.031% + 20 nA
500 pA
1.00000 mA
50 nA
0.034% + 200 nA
5 nA
10.0000 mA
500 nA
0.045% + 2 µA
50 nA
100.000 mA
5 µA
0.066% + 20 µA
500 nA
1.00000 A
1
50 µA
0.067% + 900 µA
100 µA
3.00000 A
1
500 µA
0.059% + 2.8 mA
300 µA
TEMPERATURE COEFFICIENT (0°-18°C & 28°-50°C):
±
(0.15 × accuracy specification)/
°C.
MAX. OUTPUT POWER:
110W, four quadrant source or sink operation.
SOURCE/SINK LIMITS:
±
3.15A @
±
21V,
±
1.05A @
±
105V.
CURRENT REGULATION: Line:
0.01% of range.
Load:
0.01% of range + 100
pA
.
VOLTAGE LIMIT:
Bipolar voltage limit (compliance) set with single value. Min. 0.1% of range.
OVERSHOOT:
<0.1% typical (1mA step, RL = 10k
Ω
, 20V range).
ADDITIONAL SOURCE
SPECIFICATIONS
TRANSIENT RESPONSE TIME:
30
µs minimum for the
output to recover to its spec. following a step change in load
.
COMMAND PROCESSING TIME:
Maximum time
required for the output to begin to change following the
receipt of :SOURce:VOLTage|CURRent <nrf> command.
Autorange On:
10ms.
Autorange Off:
7ms.
OUTPUT SETTLING TIME:
Time required to reach 0.1%
of final value after command is processed. 100µs typical.
Resistive load. 10µA to 100mA range, 1V step.
OUTPUT SLEW RATE (±30%):
0.25V/µs, 100V range, 100mA compliance.
0.08V/µs, 20V range, 100mA compliance.
DC FLOATING VOLTAGE:
Output can be floated up to
±250VDC from chassis ground.
REMOTE SENSE:
Up to 1V drop per load lead.
COMPLIANCE ACCURACY:
Add 0.3% of range and
±0.02% of reading to base specification.
OVER TEMPERATURE PROTECTION:
Internally
sensed temperature overload puts unit in standby mode.
RANGE CHANGE OVERSHOOT:
Overshoot into a fully
resistive 100k
Ω
load, 10Hz to 1MHz BW, adjacent range
changes between 200mV, 2V, and 20V ranges, 100mV
typical.
MINIMUM COMPLIANCE VALUE:
0.1% of range.
1. Full power source operation regardless of load to 30°C ambient. For
above 30°C and/or power sink operation, refer to the Power
Equations section of the User’s Manual.
2. For sink mode, 10µA to 100mA range, accuracy is:
±(0.5% + offset*3)
For 1A and 3A range, accuracy is:
±(1.5% + offset*3)
Содержание SourceMeter 2400
Страница 10: ......
Страница 13: ...1 PerformanceVerification...
Страница 28: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 1 16 Performance Verification...
Страница 29: ...2 Calibration...
Страница 51: ...3 RoutineMaintenance...
Страница 54: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 3 4 Routine Maintenance...
Страница 55: ...4 Troubleshooting...
Страница 68: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 4 14 Troubleshooting...
Страница 69: ...5 Disassembly...
Страница 76: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 5 8 Disassembly...
Страница 77: ...6 ReplaceableParts...
Страница 91: ...A Specifications...
Страница 111: ...B CommandReference...
Страница 123: ...C CalibrationPrograms...
Страница 130: ...C 8 Calibration Programs...
Страница 134: ......
Страница 136: ......