Calibration Programs
C-3
General program instructions
1.
With the power off, connect the Model 2400 and the digital multimeter to the IEEE-488
interface of the computer. Be sure to use shielded IEEE-488 cables for bus connections.
Also, connect the digital multimeter to the Model 2400 as discussed in Section 2.
2.
Turn on the computer, the Model 2400, and the digital multimeter. Allow the
Model 2400 and the multimeter to warm up for at least one hour before performing
calibration.
3.
Make sure the Model 2400 is set for a primary address of 24. (Use the front panel
MENU/GPIB selection to check or change the address.)
4.
Make sure the digital multimeter primary address is set to its factory default setting
(22).
5.
Make sure that the computer bus driver software (CECHP.EXE) is properly initialized.
6.
Enter the QBasic editor, and enter in the following program. Check thoroughly for
errors, and then save it using a convenient filename.
7.
Run the program and follow the prompts on the screen to perform calibration.
Program C-1. Model 2400 calibration program
' Model 2400 calibration program for use with the HP3458A Digital Multimeter.
' Rev. 1.0, 12/28/95
' 2400 primary address = 24. HP3458A primary address = 22.
OPEN "IEEE" FOR OUTPUT AS #1 ' Open IEEE-488 output path.
OPEN "IEEE" FOR INPUT AS #2 ' Open IEEE-488 input path.
PRINT #1, "INTERM CRLF" ' Set input terminator.
PRINT #1, "OUTTERM LF" ' Set output terminator.
PRINT #1, "REMOTE 22 24" ' Put 2400, 3458A in remote.
PRINT #1, "CLEAR" ' Send DCL.
PRINT #1, "OUTPUT 24;*CLS" ' Initialize 2400.
PRINT #1, "OUTPUT 24;*ESE 1;*SRE 32" ' Enable OPC and SRQ.
PRINT #1, "OUTPUT 22;PRESET NORM" ' Initialize 3458A.
PRINT #1, "OUTPUT 22;NPLC 10;NDIG 7;TRIG LINE;FUNC DCV,AUTO"
'
CLS ' Clear CRT.
PRINT "Model 2400 Calibration Program"
PRINT "Connect DMM volts input to Model 2400 INPUT/OUTPUT jacks."
GOSUB KeyCheck
FOR J = 1 TO 2 ' Select volts/current cal.
If J = 1 THEN
RESTORE VoltsInit
Max = 8
Range = .2
Limit = 200
ELSE
RESTORE CurrentInit
Max = 4
Range = .000001
Limit = 1
Содержание SourceMeter 2400
Страница 10: ......
Страница 13: ...1 PerformanceVerification...
Страница 28: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 1 16 Performance Verification...
Страница 29: ...2 Calibration...
Страница 51: ...3 RoutineMaintenance...
Страница 54: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 3 4 Routine Maintenance...
Страница 55: ...4 Troubleshooting...
Страница 68: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 4 14 Troubleshooting...
Страница 69: ...5 Disassembly...
Страница 76: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 5 8 Disassembly...
Страница 77: ...6 ReplaceableParts...
Страница 91: ...A Specifications...
Страница 111: ...B CommandReference...
Страница 123: ...C CalibrationPrograms...
Страница 130: ...C 8 Calibration Programs...
Страница 134: ......
Страница 136: ......